Pinnacle3 Radiation Therapy Planning System, Model numbers 870231 and 870237.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Model 870231: Software version 16.0.4, UDI (01)00884838099081(10)16.0.4.60330; Software version 16.4.2, UDI (01)0084838103054(10)16.4.2.60330, UDI (01)00884838102965(10)16.4.2.60330, UDI (01)00884838102972(10)16.4.2.60315, UDI (01)00884838102989(10)16.4.2.60330; UDI (01)00884838103191(10)16.4.2.60330; and Software version 16.4.3, UDI (01)00884838102965(10)16.4.3.60360. Model 870237: Software version 16.4.6, UDI (01)00884838103566(10)16.4.6.21197; and Software version 18.0.1, UDI (01)00884838103566(10)18.0.1.2149 and (01)00884838103566(10)18.0.1.21197.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Philips Medical Systems (Cleveland) Inc
- Reason for Recall:
- When computing a radiation dose in the system, the exported dose information is incorrect when there is more than one beam attached to the prescription and certain options are selected on the Edit Prescription screen.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Pinnacle3 Radiation Therapy Planning System, Model numbers 870231 and 870237.
Product Codes/Lot Numbers:
Model 870231: Software version 16.0.4, UDI (01)00884838099081(10)16.0.4.60330; Software version 16.4.2, UDI (01)0084838103054(10)16.4.2.60330, UDI (01)00884838102965(10)16.4.2.60330, UDI (01)00884838102972(10)16.4.2.60315, UDI (01)00884838102989(10)16.4.2.60330; UDI (01)00884838103191(10)16.4.2.60330; and Software version 16.4.3, UDI (01)00884838102965(10)16.4.3.60360. Model 870237: Software version 16.4.6, UDI (01)00884838103566(10)16.4.6.21197; and Software version 18.0.1, UDI (01)00884838103566(10)18.0.1.2149 and (01)00884838103566(10)18.0.1.21197.
Distribution:
Nationwide distribution
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1744-2022
Related Recalls
Due to a software issue, there is a potential image error of the Region of Interest for expansion/contraction for HFP (Head First Prone), FFS (Feet First Supine) and FFP (Feet First Prone) orientations.
Pinnacle 3 with TumorLOC, (870258) Radiation Therapy Planning System Software
Philips Medical Systems (Cleveland)
Due to software issue, Radiation Therapy Planning system may provide incorrect dataset calculations when performing the "Stopping Power Ratio" (SPR) ,
Forte Gamma Camera System is intended to produce images depicting the anatomical distribution of single-photon and positron-emitting radioisotopes within the body for interpretation by medical personnel.
Philips Medical Systems (Cleveland)
An issue with the Detector for the Forte Family of cameras may result in either detector 1 or detector 2 falling unimpeded vertically to the end stops of its travel limit.