Custom Tubing Pack with Bioline Coating UDI code: 00607567208230 Part Number: 709000058 The HLM Tubing Sets with Bioline Coating are for single use only. They may be sold sterile, non-sterile, and bulk packed. Custom tubing sets that are sold sterile are not to be re-sterilized by the user. In open heart surgery the HLM Tubing Sets with Bioline Coating are used in combination with the heart-lung machine for the oxygenation of blood and removal of carbon dioxide. The main purpose of the HLM Tubing Sets with Bioline Coating is to connect the patient to the heart-lung machine and its components. The HLM Tubing Sets with Bioline Coating are therefore a component in the extracorporeal perfusion circulation system. The utilization period of the use of the tubing sets is restricted to six hours. The Bioline Coating improves the physical surface properties of products for the extracorporeal circulation system.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Lot number 3000033848
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Datascope Corporation
- Reason for Recall:
- A complaint was received on January 1, 2017 stating the pigtail on line 2 of the oxygenator was assembled to the port where the yellow cap and the clear cap remained on the center port.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Custom Tubing Pack with Bioline Coating UDI code: 00607567208230 Part Number: 709000058 The HLM Tubing Sets with Bioline Coating are for single use only. They may be sold sterile, non-sterile, and bulk packed. Custom tubing sets that are sold sterile are not to be re-sterilized by the user. In open heart surgery the HLM Tubing Sets with Bioline Coating are used in combination with the heart-lung machine for the oxygenation of blood and removal of carbon dioxide. The main purpose of the HLM Tubing Sets with Bioline Coating is to connect the patient to the heart-lung machine and its components. The HLM Tubing Sets with Bioline Coating are therefore a component in the extracorporeal perfusion circulation system. The utilization period of the use of the tubing sets is restricted to six hours. The Bioline Coating improves the physical surface properties of products for the extracorporeal circulation system.
Product Codes/Lot Numbers:
Lot number 3000033848
Distribution:
Distributed in: US, CA
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1769-2017
Related Recalls
The Battery Charging Station is an optional dual-bay charger used to charge the Cardiosave IABP Lithium- Ion batteries when they are not being used to operate the IABP. Getinge has identified that a protruding screw in the left battery bay limits full insertion of certain batteries and prevents proper charging. The right bay of the Battery Charging Station is not affected by this issue.
The firm has identified that the battery runtime and cycle specifications contained in the devices Instructions for Use (IFU) need to be updated.