ADULT CHEST TUBE TRAY. Medical convenience kit, Avid Medical Part Number: LAKC001-04
Class I - DangerousWhat Should You Do?
- Check if you have this product: Part Number: LAKC001-04; Package GTIN: 20809160261016; Lot Number: 1404809 (Expiration Date: 08/01/2021) and 1408337 (Expiration Date: 09/01/2021) Part Number: LAKC001-04; Package GTIN: 20809160293130; Lot Number: 1427628 (Expiration Date: 09/01/2021) Part Number: LAKC001-04; Package GTIN: 20809160313005; Lot number: 1427629 (Expiration Date: 09/01/2021), 1432814 (Expiration Date: 11/30/2021), and 1436046 (Expiration Date: 11/30/2021)
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Avid Medical, Inc.
- Reason for Recall:
- Medical convenience kits are being recalled due to a potential breach in the package integrity of a component, the BD ChloraPrep 3mL applicator. Under certain circumstances the applicator may grow Aspergillus penicillioides, which may result in a breach of the package integrity. A breach of package integrity may contaminate other components in AVID Medical kits.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
ADULT CHEST TUBE TRAY. Medical convenience kit, Avid Medical Part Number: LAKC001-04
Product Codes/Lot Numbers:
Part Number: LAKC001-04; Package GTIN: 20809160261016; Lot Number: 1404809 (Expiration Date: 08/01/2021) and 1408337 (Expiration Date: 09/01/2021) Part Number: LAKC001-04; Package GTIN: 20809160293130; Lot Number: 1427628 (Expiration Date: 09/01/2021) Part Number: LAKC001-04; Package GTIN: 20809160313005; Lot number: 1427629 (Expiration Date: 09/01/2021), 1432814 (Expiration Date: 11/30/2021), and 1436046 (Expiration Date: 11/30/2021)
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1784-2021
Related Recalls
Halyard, Towel Six Pack. Kit Code: AMTS70-01.
AVID Medical
Inadvertent distribution of sealed pouches that had not undergone validated sterilization.
Halyard, Pain Pack. Kit Code: AMPK48-01.
AVID Medical
Inadvertent distribution of sealed pouches that had not undergone validated sterilization.
Halyard, Drape Pack. Kit Code: LMDP36-01.
AVID Medical
Inadvertent distribution of sealed pouches that had not undergone validated sterilization.