The ARCHITECT Estradiol Kit Intended to measure estradiol, an estrogenic steroid, in plasma.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Product List Number 07K7220 with Lot #: 55900UI01with expiration date of 04/23/2016, Lot #: 55908UI01 with expiration date of 04/29/2016, Lot #: 55941UI00 with expiration date of 04/23/2016, Lot #: 57929UI00 with expiration date of 07/16/2016, Lot #: 57931UI01 with expiration date of 07/27/2016, Lot #: 60105UI00 with expiration date of 10/02/2016 and Lot #: 61106UI00 with expiration date of 11/27/2016. Product List Number 07K7225 - Lot #: 55900UI00 with expiration date of 04/23/2016, Lot #: 55908UI00 with expiration date of 04/29/2016, Lot #: 57929UI01 with expiration date of 07/16/2016, Lot #: 57931UI00 with expiration date of 08/27/2016, Lot #: 60105UI01 with expiration date of 02/10/2016, Lot #: 61106UI01 with expiration date of 11/27/2016 and Lot #: 62263UI00 with expiration date of 11/27/2016.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Abbott Laboratories
- Reason for Recall:
- Interaction of Fulvestrant with the ARCHITECT Estradiol assay, leading to falsely elevated estradiol results in patients treated with Fulvestrant.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
The ARCHITECT Estradiol Kit Intended to measure estradiol, an estrogenic steroid, in plasma.
Product Codes/Lot Numbers:
Product List Number 07K7220 with Lot #: 55900UI01with expiration date of 04/23/2016, Lot #: 55908UI01 with expiration date of 04/29/2016, Lot #: 55941UI00 with expiration date of 04/23/2016, Lot #: 57929UI00 with expiration date of 07/16/2016, Lot #: 57931UI01 with expiration date of 07/27/2016, Lot #: 60105UI00 with expiration date of 10/02/2016 and Lot #: 61106UI00 with expiration date of 11/27/2016. Product List Number 07K7225 - Lot #: 55900UI00 with expiration date of 04/23/2016, Lot #: 55908UI00 with expiration date of 04/29/2016, Lot #: 57929UI01 with expiration date of 07/16/2016, Lot #: 57931UI00 with expiration date of 08/27/2016, Lot #: 60105UI01 with expiration date of 02/10/2016, Lot #: 61106UI01 with expiration date of 11/27/2016 and Lot #: 62263UI00 with expiration date of 11/27/2016.
Distribution:
Nationwide distribution
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1865-2016
Related Recalls
Software issue for hq analyzer results in system not visibly applying appropriate flagging to results when saturation is present which may result in incorrect results.
Brand Name: CELL-DYN Ruby Product Name: CELL-DYN Ruby Model/Catalog Number: 08H67-01 Software Version: All software versions Product Description: CELL-DYN Ruby Component: No
Abbott Laboratories
When expired reagents are scanned or manually entered, the system will change the expiration date to current or future date without notifying user that an expired reagent is being used. Use of expired reagents is against the guidelines of the Operator's Manual.
Brand Name: CELL-DYN Ruby Product Name: CELL-DYN Ruby Model/Catalog Number: 08H67-10 Software Version: All software versions Product Description: CELL-DYN Ruby Component: No
Abbott Laboratories
When expired reagents are scanned or manually entered, the system will change the expiration date to current or future date without notifying user that an expired reagent is being used. Use of expired reagents is against the guidelines of the Operator's Manual.