LAP LASER PICTOR 3D Multi Color Laser System PICTOR 3D is a LAP laser system which uses multi projection format that defines isocenters, field and MLC contours along with a 3D patient view.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Product Code: 0007569-0001
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
β οΈ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- LAP of America Laser Applications, LLC
- Reason for Recall:
- While projecting yellow lines it was notice that the yellow projection has the potential to splits into two separate red and green projections. In this case the difference between the red and green projection was up to 2.5mm.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
LAP LASER PICTOR 3D Multi Color Laser System PICTOR 3D is a LAP laser system which uses multi projection format that defines isocenters, field and MLC contours along with a 3D patient view.
Product Codes/Lot Numbers:
Product Code: 0007569-0001
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1866-2014
Related Recalls
CARINAsim MAN-1236 Rev 2 and CARINAiso MAN-1237 Rev 2 laser software. LAP isocenter marking control software. Laser systems for patient alignment.
LAP of America Laser Applications
Potential for patient to be marked incorrectly. Customers currently using CARINAsim or CARINAiso systems can encounter two safety relevant errors in the operating instructions (CARINAsim MAN-1236 Rev 2 / CARINAiso MAN-1237 Rev 2)