1) Adult Manual Resuscitator, 2K8004, Oxygen Reservoir Bag, Adult Mask; 2) Adult Manual Resuscitator, 2K8005, 40" (1.0 m) Oxygen Reservoir Tubing, Adult Mask; 3) Adult Manual Resuscitator, 2K8017, Variable Volume Oxygen Reservoir Tubing, Adult Mask; 4) Adult Manual Resuscitator, 2K8034, 40" (1.0 m) Oxygen Reservoir Tubing, PEEP Valve, without Mask; 5) Adult Manual Resuscitator, 2K8035, Oxygen Reservoir Bag, PEEP Valve, Adult Mask; 6) Pediatric Manual Resuscitator, 2K8037, 40" (1.0 m) Oxygen Reservoir Tubing, PEEP Valve, Pediatric Mask, Pressure Limiting Valve with Lock; 7) Infant Manual Resuscitator, 2K8040, Oxygen Reservoir Bag, PEEP Valve, Infant Mask, Pressure Limiting Valve with Lock Product Usage: Pulmonary resuscitation

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    1) Product Code 2K8005; Lot Numbers: 0000355656, 0000355659, 0000355662, 0000355664, 0000355665, 0000355666, 0000355668; 2) Product Code 2K8004; Lot Number: 0000355673; 3) Product Code 2K8017; Lot Numbers: 0000355676, 0000355677; 4) Product Code 2K8034; Lot Number: 0000355688; 5) Product Code 2K8035; Lot Number: 0000358023; 6) Product Code 2K8037; Lot Number: 0000358025; 7) Product Code 2K8040; Lot Number: 0000358026
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Carefusion 2200 Inc
Reason for Recall:
CareFusion has received four (4) customer reports of a component (oxygen connector) within the resuscitation bag being occluded which disables the flow of supplemental oxygen to the resuscitation bag and thus may pose a serious patient safety risk.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

1) Adult Manual Resuscitator, 2K8004, Oxygen Reservoir Bag, Adult Mask; 2) Adult Manual Resuscitator, 2K8005, 40" (1.0 m) Oxygen Reservoir Tubing, Adult Mask; 3) Adult Manual Resuscitator, 2K8017, Variable Volume Oxygen Reservoir Tubing, Adult Mask; 4) Adult Manual Resuscitator, 2K8034, 40" (1.0 m) Oxygen Reservoir Tubing, PEEP Valve, without Mask; 5) Adult Manual Resuscitator, 2K8035, Oxygen Reservoir Bag, PEEP Valve, Adult Mask; 6) Pediatric Manual Resuscitator, 2K8037, 40" (1.0 m) Oxygen Reservoir Tubing, PEEP Valve, Pediatric Mask, Pressure Limiting Valve with Lock; 7) Infant Manual Resuscitator, 2K8040, Oxygen Reservoir Bag, PEEP Valve, Infant Mask, Pressure Limiting Valve with Lock Product Usage: Pulmonary resuscitation

Product Codes/Lot Numbers:

1) Product Code 2K8005; Lot Numbers: 0000355656, 0000355659, 0000355662, 0000355664, 0000355665, 0000355666, 0000355668; 2) Product Code 2K8004; Lot Number: 0000355673; 3) Product Code 2K8017; Lot Numbers: 0000355676, 0000355677; 4) Product Code 2K8034; Lot Number: 0000355688; 5) Product Code 2K8035; Lot Number: 0000358023; 6) Product Code 2K8037; Lot Number: 0000358025; 7) Product Code 2K8040; Lot Number: 0000358026

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1947-2013