The Tina-quant lgA Gen.2 is an immunoturbidimetric assay. Anti-lgA antibodies react with antigen in the sample to form an antigen/antibody complex which is measured turbidimetrically.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Lot 368756
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Roche Diagnostics Corporation
- Reason for Recall:
- Calibration failures with Std.E, Sens.E, or Dup.E error due to atypical reaction kinetics and quality control failures have occurred due to contamination of R1 reagent with R3 reagent.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
The Tina-quant lgA Gen.2 is an immunoturbidimetric assay. Anti-lgA antibodies react with antigen in the sample to form an antigen/antibody complex which is measured turbidimetrically.
Product Codes/Lot Numbers:
Lot 368756
Distribution:
Distributed in: US, AL, IN, KY
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1979-2019
Related Recalls
cobas p 501 post-analytical units, Part Number 05158494001. For diagnostic sample storage and retrieval.
Roche Diagnostics
There is a potential for the tray input flap on the post analytical units to become loose, potentially detaching from the instrument.
cobas p 701 post-analytical units, Part Number 05349966001. For diagnostic sample storage and retrieval.
Roche Diagnostics
There is a potential for the tray input flap on the post analytical units to become loose, potentially detaching from the instrument.
Cobas¿ infinity central lab IT solution Material Number: 07154003001 Calculator/Data Processing Module, For Clinical Use
Roche Diagnostics
Using the following versions of cobas infinity software (2.0 thorough 2.5), there is a potential where the results from the first sample subsequent to the bottle change, (when using the Standby bottle function), does not display a quality control (QC) alarm when one should be present. If the QC is out of range on the stand-by bottle and it is not removed from the system, there is a potential that when the status of the Standby bottle changes from the Standby bottle to the Current bottle status, QC alarms will not be displayed next to the sample result.