OES 4000 Hysteroscope, Model/Catalog Number: A4674A
Class I - DangerousWhat Should You Do?
- Check if you have this product: Model/Catalog Number: A4674A, UDI-DI: 04042761006361, Serial Number: 805904;
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Olympus Corporation of the Americas
- Reason for Recall:
- The required leakage test was not performed following of a re-work for a non-conformity. This could potentially result in liquid ingress into the eyepiece and cause a foggy image.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
OES 4000 Hysteroscope, Model/Catalog Number: A4674A
Product Codes/Lot Numbers:
Model/Catalog Number: A4674A, UDI-DI: 04042761006361, Serial Number: 805904;
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2029-2024
Related Recalls
Brand Name: Single Use Ligating Device Product Name: Olympus HX-400U-30 Model/Catalog Number: HX-400U-30 Used with an Olympus endoscope to deliver a nylon loop snare.
Olympus Corporation of the Americas
Potential that the ligation loop was unable to release or detach as expected during use, causing the loop to become unintentionally fixed around patient anatomy.
Continued reports of positive cultures and infections have identified updates to the reprocessing materials to minimize potential deviations in TJF duodenoscope reprocessing. Olympus is informing users of these changes, reminding users to closely follow the operation and reprocessing instructions, and providing updated materials to users.
Device kits contained an incorrectly sized 4mm x 4cm urinary tract balloon catheter (BURS0404) instead of the correct 8mm x 15cm nephrostomy tract balloon catheter (BPCN0815). All other components in the kits were correctly included and packaged in the kit.