Brand Name: VITROS Product Name: VITROS 4600 Chemistry System Model/Catalog Number: (1) 6802445, (2) 6900440 Software Version: version 3.2 to version 3.8.3 (all available software versions) Product Description: In vitro diagnostic chemistry analyzer.
Class I - DangerousWhat Should You Do?
- Check if you have this product: VITROS 4600 Chemistry System; Product Code: 6802445; UDI: 10758750012343; VITROS 4600 Chemistry System - Refurbished; Product Code: 6900440; UDI: 10758750033201; Range of installed serial numbers:46000111-46001897; Manufacture date range: April 2011 to present Distribution date range; April 2011 to present Total systems currently installed in the US: 135 Total systems currently installed outside the US: 1,127 Total systems installed worldwide (Product Quantity Distributed): 1,262 Note1: The catalogue number 6802445 is used to denote the VITROS 4600 Chemistry System. Catalogue number 6900440 is used to denote VITROS 4600 Chemistry System- refurbished. A VITROS 4600 Chemistry System maintains its original serial number through the refurbishment process. Therefore, the manufacturing and distribution information for the two catalogue numbers will be combined. Note2: When manufactured, the system s serial numbers are sequential, but not all analyzers are released for distribution and therefore not installed. The range of serial numbers provided includes the identified minimum and maximum in serial numbers associated with all currently installed systems. Therefore, the amount of each system manufactured may not equal the amount of each system installed. Additionally, not all serial numbers listed are associated with an active installed analyzer in use today. Note3: The value given for "Product Quantity Distributed" is the total number of active installed systems that are in use today. This number is known to fluctuate as individual systems are installed and/or decommissioned.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Ortho-Clinical Diagnostics, Inc.
- Reason for Recall:
- a software anomaly allows test results to be reported using Micro Tip and Micro Slide diluent packs that the system treats as expired even when the Use Expired Reagents setting is not enabled if a shelf expiration date is omitted during manual loading. The issue was identified by a customer complaint, in which a CREA Urine test was processed with an RE (Reagent Expired) code, even though the diluent pack was not expired and the use-expired-reagents setting was disabled. This can result in erroneous test results and/or delaying test results.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Brand Name: VITROS Product Name: VITROS 4600 Chemistry System Model/Catalog Number: (1) 6802445, (2) 6900440 Software Version: version 3.2 to version 3.8.3 (all available software versions) Product Description: In vitro diagnostic chemistry analyzer.
Product Codes/Lot Numbers:
VITROS 4600 Chemistry System; Product Code: 6802445; UDI: 10758750012343; VITROS 4600 Chemistry System - Refurbished; Product Code: 6900440; UDI: 10758750033201; Range of installed serial numbers:46000111-46001897; Manufacture date range: April 2011 to present Distribution date range; April 2011 to present Total systems currently installed in the US: 135 Total systems currently installed outside the US: 1,127 Total systems installed worldwide (Product Quantity Distributed): 1,262 Note1: The catalogue number 6802445 is used to denote the VITROS 4600 Chemistry System. Catalogue number 6900440 is used to denote VITROS 4600 Chemistry System- refurbished. A VITROS 4600 Chemistry System maintains its original serial number through the refurbishment process. Therefore, the manufacturing and distribution information for the two catalogue numbers will be combined. Note2: When manufactured, the system s serial numbers are sequential, but not all analyzers are released for distribution and therefore not installed. The range of serial numbers provided includes the identified minimum and maximum in serial numbers associated with all currently installed systems. Therefore, the amount of each system manufactured may not equal the amount of each system installed. Additionally, not all serial numbers listed are associated with an active installed analyzer in use today. Note3: The value given for "Product Quantity Distributed" is the total number of active installed systems that are in use today. This number is known to fluctuate as individual systems are installed and/or decommissioned.
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2040-2025
Related Recalls
VITROS Chemistry Products LAC Slides. Catalog Number: 8150112. 1 Sales Unit = contains 5 cartridges of 18 slides, for a total of 90 tests.
Ortho-Clinical Diagnostics
VITROS Chemistry Products LAC Slides on VITROS XT 3400 and VITROS XT 7600 Systems are experiencing an increased occurrence of calibration failure related to lots from coating 0130 and above with Condition Code TH4-63J during calibration, causing the delay of lac test results and potentially impact patient management. The issue was identified by QuidelOrtho through an increase in customer complaints compared to the historical baseline with average failure rate of 1% for coating 0130 in 2025 while average of failure rate for other coating was 0.06% in 2025 and average of failure rate for all coatings is 0.03% for 2024.
Chemistry Products LAC Slides. Catalog Number: 8433880. 1 Sales Unit = contains 5 cartridges of 60 slides, for a total of 300 tests.
Ortho-Clinical Diagnostics
VITROS Chemistry Products LAC Slides on VITROS XT 3400 and VITROS XT 7600 Systems are experiencing an increased occurrence of calibration failure related to lots from coating 0130 and above with Condition Code TH4-63J during calibration, causing the delay of lac test results and potentially impact patient management. The issue was identified by QuidelOrtho through an increase in customer complaints compared to the historical baseline with average failure rate of 1% for coating 0130 in 2025 while average of failure rate for other coating was 0.06% in 2025 and average of failure rate for all coatings is 0.03% for 2024.
VITROS Chemistry Products Ca Slides; Catalog Number: 145 0261;
Ortho-Clinical Diagnostics
Generations 67 and above of VITROS Ca Slides are experiencing an increased occurrence of Condition Code TH4-63J during calibration on VITROS XT 3400 and VITROS XT 7600 Systems, causing calibration failures that delay calcium test results and potentially impact patient management. The imaging reflectometer algorithm detects excessive "spikes" (imperfections) in slide images and generates the TH4-63J condition code when the threshold of 600 spikes is exceeded. Updated analysis revealed a 19-fold increase in failure rates from 0.0043% (2024 baseline) to 0.082% for Generation 67+ slides in 2025, with 100% of calibrator failures occurring at the lowest concentration level. This can result in delayed diagnosis and treatment of conditions such as hypercalcemic crisis, parathyroid disorders, and other calcium-related emergencies requiring urgent intervention.