RUSCH, TruLite, One-Piece Disposable Laryngoscope Handle Blade Combination, Non-Sterile, Rx Only, Distributed by: Teleflex Medical.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Product Code: 004660000, (Size Miller 0), Lot numbers: 1306602, 1308602 & 1310602; Product Code: 004660001, (Size Miller 1), Lot numbers: 1306612, 1308612 & 1310612; Product Code: 004660002, (Size Miller 2), Lot numbers: 1306622, 1308622, 1309622 & 1310622; Product Code: 004660003, (Size Miller 3), Lot numbers: 1306632 & 1310632; Product Code: 004660004, (Size Miller 4), Lot numbers: 1306642, 1308642 & 1309642; Product Code: 004660010, (Size Miller 00), Lot numbers: 1306592, 1308592, 1309592, 1312592 & 1401592; Product Code: 004661002, (Size Mac 2), Lot numbers: 1306522, 1308522 & 1310522; Product Code: 004661003, (Size Mac 3), Lot numbers: 1306532, 1308532 & 1310532 and Product Code: 004661004, (Size Mac 4), Lot numbers: 306542, 1308542 & 1309542.
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Teleflex Medical
Reason for Recall:
There is a possibility the handle of the laryngoscope may heat up and melt the end cap.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

RUSCH, TruLite, One-Piece Disposable Laryngoscope Handle Blade Combination, Non-Sterile, Rx Only, Distributed by: Teleflex Medical.

Product Codes/Lot Numbers:

Product Code: 004660000, (Size Miller 0), Lot numbers: 1306602, 1308602 & 1310602; Product Code: 004660001, (Size Miller 1), Lot numbers: 1306612, 1308612 & 1310612; Product Code: 004660002, (Size Miller 2), Lot numbers: 1306622, 1308622, 1309622 & 1310622; Product Code: 004660003, (Size Miller 3), Lot numbers: 1306632 & 1310632; Product Code: 004660004, (Size Miller 4), Lot numbers: 1306642, 1308642 & 1309642; Product Code: 004660010, (Size Miller 00), Lot numbers: 1306592, 1308592, 1309592, 1312592 & 1401592; Product Code: 004661002, (Size Mac 2), Lot numbers: 1306522, 1308522 & 1310522; Product Code: 004661003, (Size Mac 3), Lot numbers: 1306532, 1308532 & 1310532 and Product Code: 004661004, (Size Mac 4), Lot numbers: 306542, 1308542 & 1309542.

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-2130-2014