da Vinci SP Camera; PN 430060-33; Model Number: 430060-33. For endoscopic surgery.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Model Number: 430060-33; UDI Number: 00886874114384; All serial numbers.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Intuitive Surgical, Inc.
- Reason for Recall:
- Some cameras were found to have residual soil adhered to the tip of the camera after reprocessing which may affect image quality resulting in needing to switch to a second camera or converting to another surgical modality. If residual soil makes contact with the patient, exposure could result in sensitization making future component blood therapy or tissue typing more challenging; additionally if a small piece of inert epoxy were to be retained, scar tissue could form.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
da Vinci SP Camera; PN 430060-33; Model Number: 430060-33. For endoscopic surgery.
Product Codes/Lot Numbers:
Model Number: 430060-33; UDI Number: 00886874114384; All serial numbers.
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2131-2020
Related Recalls
Due to an increase in complaints concerning foot tray pedal spring failing resulting in the pedal remaining pressed
8MM,HORIZON SMALL CLIP APPLIER,IS4000¿ REF 470401
Intuitive Surgical
Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws.
8MM,MEGA SUTURECUT ND,IS4000 REF 471309
Intuitive Surgical
Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws.