KODAK DirectView DR 9000 System, Catalog Numbers 1966688 (US and Canada) and 8161937 (outside the US and Canada) --- A permanently installed diagnostic x-ray system composed of 3 main components: an operator console, U-Arm assembly with x-ray tube including a collimator, and a stationary generator.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Service Code 1534
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Carestream Health Inc.
Reason for Recall:
A control system failure may cause the U-Arm to move downward unexpectedly when the Radiology Technologist is positioning the equipment using the "Up" or "Down" control buttons. Downward movement occurs at three times the normal rate of speed and will continue until the device control button is released, the Emergency Stop button is activated or the equipment Bump Sensor contacts an object.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

KODAK DirectView DR 9000 System, Catalog Numbers 1966688 (US and Canada) and 8161937 (outside the US and Canada) --- A permanently installed diagnostic x-ray system composed of 3 main components: an operator console, U-Arm assembly with x-ray tube including a collimator, and a stationary generator.

Product Codes/Lot Numbers:

Service Code 1534

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-2137-2014

Related Recalls

There is a potential for unexpected failure of the electrical components within the Carestream Health Inc. generator installed in your DRX-Revolution Mobile X-ray System(s). This can lead to temporary and self contained thermal overload within the generator. Should an event like this occur, the system will become inoperable and loud noise, burnt smell and smoke may be detectable.

Oct 2, 2023 Diagnostic Equipment Nationwide View Details →

After pressing and releasing the Z-Axis Motorized buttons on the Tube Head Display, the Overhead Tube Crane (OTC) could potentially continue the movement unexpectedly to a certain distance which could potentially result in a injury

Jan 30, 2023 Diagnostic Equipment Nationwide View Details →