Target Detachable Coils: TARGET HELICAL NANO 1.5 MM X 2 CM MODEL Number:M0035431520 Neurology: Target Detachable Coils are intended to endovascularly obstruct or occlude blood flow in vascular abnormalities of the neurovascular and peripheral vessels.
Class I - DangerousWhat Should You Do?
- Check if you have this product: MODEL Number: M0035431520: Lot Number/Expiration Date: 18784673 31-Jan-19 18723226 30-Nov-18 18686302 30-Nov-18 18637332 31-Oct-18 18627687 31-Oct-18 18569824 31-Aug-18 18453278 30-Jun-18 18391505 31-May-18 18769932 31-Dec-18 18735994 30-Nov-18 18710356 30-Nov-18 18652851 31-Oct-18 18801847 31-Jan-19 18755945 31-Dec-18 18602481 30-Sep-18 18832966 28-Feb-19 18843005 28-Feb-19 18630099 31-Oct-18 18510994 31-Jul-18 18495363 31-Jul-18 18484141 31-Jul-18 18803162 31-Jan-19 18786297 31-Jan-19 18754879 31-Dec-18 18734778 30-Nov-18 18602423 30-Sep-18 18585290 31-Aug-18 18450430 30-Jun-18 18769816 31-Dec-18 18785485 31-Jan-19 18637274 31-Oct-18 18511110 31-Jul-18 18465901 30-Jun-18 18769700 31-Dec-18 18746995 31-Dec-18 18725103 30-Nov-18 18694228 30-Nov-18 18686012 30-Nov-18 18665561 31-Oct-18 18607752 30-Sep-18 18449575 30-Jun-18 18832355 28-Feb-19 18831744 31-Jan-19 18842138 28-Feb-19 18732878 30-Nov-18 18637216 31-Oct-18 18584252 31-Aug-18 18569882 31-Aug-18 18569766 31-Aug-18 18523410 31-Jul-18 19033487 30-Apr-19 18984845 30-Apr-19 18972140 30-Apr-19 18861952 28-Feb-19 18878607 28-Feb-19 19063170 31-May-19 18923191 31-Mar-19 18897527 28-Feb-19 18897185 28-Feb-19 18910290 31-Mar-19 19032199 30-Apr-19 19002213 30-Apr-19 19002155 30-Apr-19 18950942 31-Mar-19 18897698 28-Feb-19 18858983 28-Feb-19 19031329 30-Apr-19 19031909 30-Apr-19 19002271 30-Apr-19 18951058 31-Mar-19 18950884 31-Mar-19 18896843 28-Feb-19 18897014 28-Feb-19 18879010 28-Feb-19 18909942 31-Mar-19 18910638 31-Mar-19 18843872 28-Feb-19 19031619 30-Apr-19 19002329 30-Apr-19 18878204 28-Feb-19 18886556 28-Feb-19 18938172 31-Mar-19 18939594 31-Mar-19 18951000 31-Mar-19 18897356 28-Feb-19 18885142 28-Feb-19 18971589 30-Apr-19 18950826 31-Mar-19 18860449 28-Feb-19.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Stryker Neurovascular
- Reason for Recall:
- Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended stretch resistance. The sutures inside the coil that provide stretch resistance may have been damaged during manufacturing which can lead to an increased occurrence of coil stretching.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Target Detachable Coils: TARGET HELICAL NANO 1.5 MM X 2 CM MODEL Number:M0035431520 Neurology: Target Detachable Coils are intended to endovascularly obstruct or occlude blood flow in vascular abnormalities of the neurovascular and peripheral vessels.
Product Codes/Lot Numbers:
MODEL Number: M0035431520: Lot Number/Expiration Date: 18784673 31-Jan-19 18723226 30-Nov-18 18686302 30-Nov-18 18637332 31-Oct-18 18627687 31-Oct-18 18569824 31-Aug-18 18453278 30-Jun-18 18391505 31-May-18 18769932 31-Dec-18 18735994 30-Nov-18 18710356 30-Nov-18 18652851 31-Oct-18 18801847 31-Jan-19 18755945 31-Dec-18 18602481 30-Sep-18 18832966 28-Feb-19 18843005 28-Feb-19 18630099 31-Oct-18 18510994 31-Jul-18 18495363 31-Jul-18 18484141 31-Jul-18 18803162 31-Jan-19 18786297 31-Jan-19 18754879 31-Dec-18 18734778 30-Nov-18 18602423 30-Sep-18 18585290 31-Aug-18 18450430 30-Jun-18 18769816 31-Dec-18 18785485 31-Jan-19 18637274 31-Oct-18 18511110 31-Jul-18 18465901 30-Jun-18 18769700 31-Dec-18 18746995 31-Dec-18 18725103 30-Nov-18 18694228 30-Nov-18 18686012 30-Nov-18 18665561 31-Oct-18 18607752 30-Sep-18 18449575 30-Jun-18 18832355 28-Feb-19 18831744 31-Jan-19 18842138 28-Feb-19 18732878 30-Nov-18 18637216 31-Oct-18 18584252 31-Aug-18 18569882 31-Aug-18 18569766 31-Aug-18 18523410 31-Jul-18 19033487 30-Apr-19 18984845 30-Apr-19 18972140 30-Apr-19 18861952 28-Feb-19 18878607 28-Feb-19 19063170 31-May-19 18923191 31-Mar-19 18897527 28-Feb-19 18897185 28-Feb-19 18910290 31-Mar-19 19032199 30-Apr-19 19002213 30-Apr-19 19002155 30-Apr-19 18950942 31-Mar-19 18897698 28-Feb-19 18858983 28-Feb-19 19031329 30-Apr-19 19031909 30-Apr-19 19002271 30-Apr-19 18951058 31-Mar-19 18950884 31-Mar-19 18896843 28-Feb-19 18897014 28-Feb-19 18879010 28-Feb-19 18909942 31-Mar-19 18910638 31-Mar-19 18843872 28-Feb-19 19031619 30-Apr-19 19002329 30-Apr-19 18878204 28-Feb-19 18886556 28-Feb-19 18938172 31-Mar-19 18939594 31-Mar-19 18951000 31-Mar-19 18897356 28-Feb-19 18885142 28-Feb-19 18971589 30-Apr-19 18950826 31-Mar-19 18860449 28-Feb-19.
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2169-2016
Related Recalls
Stryker Neurovascular has observed that devices with Out of United States (OUS) approved Instructions for Use (IFU) and Packaging Labels (pouch and carton), different configurations, have been erroneously distributed to customers in the US.
Stryker Neurovascular has observed an increased frequency of PTFE coating damage occurring on the Synchro Guidewires that may be caused by the practice of backloading the guidewire through the optional introducer accessory. This issue is limited to certain lots of the Synchro Guidewire that contain an older version of the introducer accessory. Due to variation in the manufacturing process of the supplier of the introducer accessory, certain lots of introducers have sharper than intended edges that can peel off the PTFE coating when physicians use a technique known as backloading. Users with impacted product in their inventory are cautioned not to use this backloading technique.
Stryker Neurovascular has observed an increased frequency of PTFE coating damage occurring on the Synchro Guidewires that may be caused by the practice of backloading the guidewire through the optional introducer accessory. This issue is limited to certain lots of the Synchro Guidewire that contain an older version of the introducer accessory. Due to variation in the manufacturing process of the supplier of the introducer accessory, certain lots of introducers have sharper than intended edges that can peel off the PTFE coating when physicians use a technique known as backloading. Users with impacted product in their inventory are cautioned not to use this backloading technique.