Ingenia 1.5T S Model Number (REF): 781347;
Class I - DangerousWhat Should You Do?
- Check if you have this product: Model Number (REF): 781347; UDI-DI: 00884838068421; Serial Numbers: 81001, 81003, 81004, 81005, 81006, 81007, 81008, 81009, 81010, 81011, 81013, 81014, 81015, 81016, 81017, 81018, 81019, 81020, 81023, 81024, 81025, 81027, 81028, 81029, 81030, 81031, 81032, 81033, 81034, 81035, 81037, 81038, 81040, 81041, 81042, 81043, 81044, 81045, 81046, 81047, 81048, 81049, 81050, 81051, 81052, 81055, 81056, 81058, 81059, 81060, 81061, 81062, 81063, 81064, 81065, 81067, 81070, 81071, 81072, 81074, 81075, 81076, 81077, 81078, 81079, 81080, 81084, 81085, 81086, 81087, 81088, 81090, 81091, 81092, 81093, 81094, 81095, 81096, 81097, 81098, 81103, 81105, 81106, 81108, 81111, 81112, 81113, 81114, 81115, 81116, 81117, 81118, 81122, 81123, 81124, 81126, 81127, 81128, 81129, 81131, 81132, 81133, 81137, 81138, 81139, 81140, 81145, 81146, 81147, 81148, 81149, 81151, 81152, 81153, 81154, 81155, 81156, 81157, 81158, 81159, 81162, 81164, 81165, 81167, 81170, 81171, 81172, 81173, 81174, 81175, 81176, 81178, 81179, 81180, 81184, 81185;
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Philips North America
- Reason for Recall:
- The potential for component failures in the Gradient Coil of the affected MR systems may act as a heat source with a potential to produce smoke and/or fire.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Ingenia 1.5T S Model Number (REF): 781347;
Product Codes/Lot Numbers:
Model Number (REF): 781347; UDI-DI: 00884838068421; Serial Numbers: 81001, 81003, 81004, 81005, 81006, 81007, 81008, 81009, 81010, 81011, 81013, 81014, 81015, 81016, 81017, 81018, 81019, 81020, 81023, 81024, 81025, 81027, 81028, 81029, 81030, 81031, 81032, 81033, 81034, 81035, 81037, 81038, 81040, 81041, 81042, 81043, 81044, 81045, 81046, 81047, 81048, 81049, 81050, 81051, 81052, 81055, 81056, 81058, 81059, 81060, 81061, 81062, 81063, 81064, 81065, 81067, 81070, 81071, 81072, 81074, 81075, 81076, 81077, 81078, 81079, 81080, 81084, 81085, 81086, 81087, 81088, 81090, 81091, 81092, 81093, 81094, 81095, 81096, 81097, 81098, 81103, 81105, 81106, 81108, 81111, 81112, 81113, 81114, 81115, 81116, 81117, 81118, 81122, 81123, 81124, 81126, 81127, 81128, 81129, 81131, 81132, 81133, 81137, 81138, 81139, 81140, 81145, 81146, 81147, 81148, 81149, 81151, 81152, 81153, 81154, 81155, 81156, 81157, 81158, 81159, 81162, 81164, 81165, 81167, 81170, 81171, 81172, 81173, 81174, 81175, 81176, 81178, 81179, 81180, 81184, 81185;
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2205-2025
Related Recalls
Azurion R3.0. Model Number 722234. Commercial Name: Azurion 7 M20. INTERVENTIONAL FLUOROSCOPIC XRAY SYSTEM
Philips North America
Between July 2024 and June 2025, Philips Azurion R3.0 systems with a FlexArm stand were shipped without the Source-to-Skin Distance (SSD) Spacer. The SSD spacer is necessary to meet the FDA minimum source-skin distance requirement of 38 cm for standard fluoroscopy applications.
IntelliVue MP90. Product Number: M8010A.
Philips North America
Potential issue where the IntelliVue monitors did not alarm.
IntelliVue MP30. Product Number: M8002A.
Philips North America
Potential issue where the IntelliVue monitors did not alarm.