American Science & Engineering Inc,Gemini Cabinet X-ray system.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Gemini systems
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
โ ๏ธ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- American Science & Engineering, Inc.
- Reason for Recall:
- During factory testing, we discovered that Gemini systems could, in rare circumstances, experience a fault that leads to a state in which X-rays are being produced but the X-ray indicator lights (located on the operator's console and on both ends of the tunnel) are not illuminated.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
American Science & Engineering Inc,Gemini Cabinet X-ray system.
Product Codes/Lot Numbers:
Gemini systems
Distribution:
Distributed in: US, NY, FL, TN, PA, NJ, DE, ME, TX, CT, MS, NC, MD, MO, VA, DC, NV, GA
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2206-2015