ProxiDiagnost N90, UDI 706100
Class I - DangerousWhat Should You Do?
- Check if you have this product: Serial Numbers: SN18000022 SN19000005 10000047 SN18000019 20000047 20000054 10000011 10000021 10000010 10000026 10000015 SN18000011 10000000 10000009 10000004 10000005 SN18000015 SN19000008 SN19000009 10000019 10000036 20000052 SN18000007 10000033 SN18000014 10000001 10000039 20000051 SN18000004 SN18000016 SN18000018 SN19000004 SN19000011 SN19000015 SN19000018 10000008 20000045 20000053 10000006 10000022 10000048 10000017 10000012 20000039 20000044 SN19000014 10000040 10000043 SN19000017 20000026 SN19000012 SN18000006 SN18000012 SN18000017 10000014 SN18000013 SN19000013 10000030 SN18000225 SN18000024 10000049 SN19000007 SN19000010 10000041 10000046 SN18000009 SN19000001 SN19000016 10000023 10000024 10000025 10000050 20000043 10000028 SN18000002 10000020 SN18000003 SN18000023 SN19000002 SN19000003 SN19000006 10000013 10000016 10000018 SN18000020 SN18000005 20000048
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Philips North America, LLC
- Reason for Recall:
- The thermo switches in the three-phase transformer, which normally aid in powering down the System Power Distribution Unit, may be incorrectly installed and not working. If the transformer overheats due to a first failure and the thermo switch is not activated, the device may begin to generate smoke.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
ProxiDiagnost N90, UDI 706100
Product Codes/Lot Numbers:
Serial Numbers: SN18000022 SN19000005 10000047 SN18000019 20000047 20000054 10000011 10000021 10000010 10000026 10000015 SN18000011 10000000 10000009 10000004 10000005 SN18000015 SN19000008 SN19000009 10000019 10000036 20000052 SN18000007 10000033 SN18000014 10000001 10000039 20000051 SN18000004 SN18000016 SN18000018 SN19000004 SN19000011 SN19000015 SN19000018 10000008 20000045 20000053 10000006 10000022 10000048 10000017 10000012 20000039 20000044 SN19000014 10000040 10000043 SN19000017 20000026 SN19000012 SN18000006 SN18000012 SN18000017 10000014 SN18000013 SN19000013 10000030 SN18000225 SN18000024 10000049 SN19000007 SN19000010 10000041 10000046 SN18000009 SN19000001 SN19000016 10000023 10000024 10000025 10000050 20000043 10000028 SN18000002 10000020 SN18000003 SN18000023 SN19000002 SN19000003 SN19000006 10000013 10000016 10000018 SN18000020 SN18000005 20000048
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2256-2020
Related Recalls
Philips SmartPath to dStream for 3.0T with MR Elastography (MRE). 1. Model Number (REF): 782145.
Philips North America
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
Philips SmartPath to Ingenia Elition X with MR Elastography (MRE). 1. Model Number (REF): 782118. 2. Model Number (REF): 782144. 3. Model Number (REF): 782163.
Philips North America
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.