Field Upgrade Kits: P/N 5451208 RIU, 9 INCH, TOSHIBA; P/N 5485304 SERVICE PART, IMAGE INTENSIFIER, 9 INCH, TOSHIBA; P/N 5761613 SERVICE PART, IMAGE INTENSIFIER, 9 INCH - used with OEC Flexiview 8800, OEC 9800, and OEC 9900 Systems
Class I - DangerousWhat Should You Do?
- Check if you have this product: P/N/Sales Order Number: 5451208/80165102, 80171955, 80173514, 80162453, 80183950, 80151733, 80163547, 80171532, 80156410, 80176650, 10530514, 105449333, 80167595; 5485304/10557644, 118687327; 5761613/10283014, 10152211, 105108003, 105114645. OEC 9800, UDI-DI: 00840682114349; OEC 9900 Systems, UDI-DI: 00840682114301
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- GE OEC Medical Systems, Inc
- Reason for Recall:
- Image intensifiers on systems, used to provide fluoroscopic/digital spot images can become detached if the systems encounter a large impact force when moved, which may result in the image intensifier falling, which could lead to injuries such as fractures, hematoma, contusion, bruise, pain.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Field Upgrade Kits: P/N 5451208 RIU, 9 INCH, TOSHIBA; P/N 5485304 SERVICE PART, IMAGE INTENSIFIER, 9 INCH, TOSHIBA; P/N 5761613 SERVICE PART, IMAGE INTENSIFIER, 9 INCH - used with OEC Flexiview 8800, OEC 9800, and OEC 9900 Systems
Product Codes/Lot Numbers:
P/N/Sales Order Number: 5451208/80165102, 80171955, 80173514, 80162453, 80183950, 80151733, 80163547, 80171532, 80156410, 80176650, 10530514, 105449333, 80167595; 5485304/10557644, 118687327; 5761613/10283014, 10152211, 105108003, 105114645. OEC 9800, UDI-DI: 00840682114349; OEC 9900 Systems, UDI-DI: 00840682114301
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2279-2023
Related Recalls