Siemens Luminos dRF, Ysio or Uroskop Omnia systems. The Axiom Luminos dRF is intended to be used as a universal diagnostic imaging system for radiographic and fluoroscopic studies.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    model numbers 10094200, 10281163, 10281013, 10094910 with multiple serial numbers
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Siemens Medical Solutions USA, Inc
Reason for Recall:
It was discovered that during a RAD examination using Siemens Luminos dRF, Ysio or Uroskop Omnia systems with software version VB10C to VB10F and automatic exposure control, a highly unlikely malfunction can result in longer radiation exposure than required. This may also result in an overexposed image that is not of diagnostic quality. As a result the taken examination needs to be repeated. Thi
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Siemens Luminos dRF, Ysio or Uroskop Omnia systems. The Axiom Luminos dRF is intended to be used as a universal diagnostic imaging system for radiographic and fluoroscopic studies.

Product Codes/Lot Numbers:

model numbers 10094200, 10281163, 10281013, 10094910 with multiple serial numbers

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-2292-2014

Related Recalls

During patient examinations, the system may sporadically display a 0 x-ray dose administered to the patient. This behavior has only been observed during fluoroscopy and/or single-frame acquisitions. Despite the reported x-ray dose showing 0, an actual x-ray dose is applied to the patient.

Mar 9, 2026 Other Medical Devices View Details β†’
Class I - Dangerous

During patient examinations, the system may sporadically display a 0 x-ray dose administered to the patient. This behavior has only been observed during fluoroscopy and/or single-frame acquisitions. Despite the reported x-ray dose showing 0, an actual x-ray dose is applied to the patient.

Mar 9, 2026 Other Medical Devices View Details β†’

ARTIS Pheno VE30A and VE40A, Model 10849000

Siemens Medical Solutions USA

Class I - Dangerous

During 3D acquisitions, lighter and darker patient images may be captured, which may result in less accurate 3D reconstruction. The variation of brightness is a result of a constant unregulated medium dose and the angular change of patient diameter caused by rotational acquisition. This may result in unintentional low-dose radiation exposure to a patient.

Mar 12, 2026 Other Medical Devices View Details β†’