Babytherm 8004/8010, Infant Warmer, Class II device, 510 K K971198, Catalog # FR00105 (Babytherm 8010), 2M30404 (Babytherm 8004), non sterile, software controlled. Product Usage: Babytherm 8004/8010 is an open care infant warmer system for warming premature babies, neonates and infants with a body weight of up to 8 kg (17.6 lbs). The device is intended for use in delivery rooms, operating rooms, maternity and obstetric units, neonatal and pediatric wards, neonatal and pediatric intensive care units. Therapy and nursing uses: - As a resuscitation table and in birthing rooms, obstetric units, and NICUs - As infant warmer bassinet during normal and intensive care - For infant re-warming and compensation of heat loss - For thermal stabilizing of infants with mattress heater (Babytherm 8010 only) - For gentle warming therapy using the combination of radiant and mattress heater - Cooling infants with fever (temperature of mattress below infant body temperature, Babytherm 8010 only) - With adjustable bed tilt for treatment of infants in Trendelenburg and reverse Trendelenburg position - For weaning infants from incubators - For O2 and nebulizer therapy (with appropriate accessories) - For lowering an infant s bilirubin levels when using the phototherapy option
Class I - DangerousWhat Should You Do?
- Check if you have this product: Serial Numbers: ARLN-xxxx to ASFE-xxxx Catalog Numbers: FR00105 (Babytherm 8010), 2M30404 (Babytherm 8004)
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Draeger Medical, Inc.
- Reason for Recall:
- Risk associated with operating the device in skin temperature mode with temperature sensors attached but no patient connected.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Babytherm 8004/8010, Infant Warmer, Class II device, 510 K K971198, Catalog # FR00105 (Babytherm 8010), 2M30404 (Babytherm 8004), non sterile, software controlled. Product Usage: Babytherm 8004/8010 is an open care infant warmer system for warming premature babies, neonates and infants with a body weight of up to 8 kg (17.6 lbs). The device is intended for use in delivery rooms, operating rooms, maternity and obstetric units, neonatal and pediatric wards, neonatal and pediatric intensive care units. Therapy and nursing uses: - As a resuscitation table and in birthing rooms, obstetric units, and NICUs - As infant warmer bassinet during normal and intensive care - For infant re-warming and compensation of heat loss - For thermal stabilizing of infants with mattress heater (Babytherm 8010 only) - For gentle warming therapy using the combination of radiant and mattress heater - Cooling infants with fever (temperature of mattress below infant body temperature, Babytherm 8010 only) - With adjustable bed tilt for treatment of infants in Trendelenburg and reverse Trendelenburg position - For weaning infants from incubators - For O2 and nebulizer therapy (with appropriate accessories) - For lowering an infant s bilirubin levels when using the phototherapy option
Product Codes/Lot Numbers:
Serial Numbers: ARLN-xxxx to ASFE-xxxx Catalog Numbers: FR00105 (Babytherm 8010), 2M30404 (Babytherm 8004)
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2295-2016
Related Recalls
The Atlan anesthesia workstation either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use.
The Atlan anesthesia workstation either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use.
Brand Name: Atlan Product Name: Atlan A350XL Model/Catalog Number: 8621600 Software Version: Not Applicable Component: Not applicable
Draeger Medical
The Atlan anesthesia workstation either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use.