Alere INRatio PT/INR System Professional Model Numbers: 0100004 Product Usage: Alere INRatio ¿ 2 PT/INR Monitoring System (Professional Use): The Alere INRatio ¿ 2 PT/INR Monitoring System (Professional Use), consisting of the INRatio ¿ 2 Monitor and INRatio¿ 2 PT/INR test strip, is used for quantitative determination of international normalized ratio (INR) in fresh capillary whole blood to monitor the effect of warfarin on clotting time by health care professionals. The Alere INRatio¿ 2 PT/INR Monitoring System (Professional Use) is intended for use outside of the body (in vitro diagnostic use). The Alere INRatio¿ 2 PT/INR Monitoring System (Professional Use) is not intended to be used for screening purposes.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    All Lots All Serial Numbers
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Alere San Diego, Inc.
Reason for Recall:
Alere San Diego is initiating a voluntary recall for the Alere INRatio¿/INRatio¿2 PT/INR Monitoring System (Test Strips and Meters). This action is being initiated as a result of patients who had a therapeutic or near-therapeutic INR with the Alere INRatio¿/INRatio¿2 PT/INR Monitoring System but a significantly higher INR (outside of therapeutic range) when performed by a central laboratory.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Alere INRatio PT/INR System Professional Model Numbers: 0100004 Product Usage: Alere INRatio ¿ 2 PT/INR Monitoring System (Professional Use): The Alere INRatio ¿ 2 PT/INR Monitoring System (Professional Use), consisting of the INRatio ¿ 2 Monitor and INRatio¿ 2 PT/INR test strip, is used for quantitative determination of international normalized ratio (INR) in fresh capillary whole blood to monitor the effect of warfarin on clotting time by health care professionals. The Alere INRatio¿ 2 PT/INR Monitoring System (Professional Use) is intended for use outside of the body (in vitro diagnostic use). The Alere INRatio¿ 2 PT/INR Monitoring System (Professional Use) is not intended to be used for screening purposes.

Product Codes/Lot Numbers:

All Lots All Serial Numbers

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-2354-2016

Related Recalls

Cholesterol test cassettes package insert is to be updated with new bilirubin interference information because it was found that samples containing bilirubin, at concentrations greater than 2.0 mg/dL (conjugated) or 1.6 mg/dL (unconjugated), can result in low recovery of Triglyceride (TRG) and High-Density Lipoproteins (HDL).

Jun 24, 2022 Diagnostic Equipment Nationwide View Details →

The polarity of the output of power supply was reversed. Due to polarity of the output of power supplies being reversed, it renders the power supplies nonfunctional and the analyzer inoperable.

Feb 15, 2019 Diagnostic Equipment Nationwide View Details →