Instructions for Use (IFU) and Manual NobPrl CC Guided Surgery Kit Instructions for Use (IFU): "NobelGuide for NobelParallel Conical Connection", id # 2002 and Manual: Guided Surgery with NobelParallel Conical Connection", id # 79423. Article # 38072, Nobel Parallel CC Guided Surgery Kit. The Guided Surgery Kit is used as part of the installation process of NobelParallel CC Implants. The purpose of the IFU is to instruct the user how to use the products safely. The purpose of the manual is to guide the user through the usage of NobelGuide for NobelParallel CC.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Article 38072-all lots
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Nobel Biocare Usa Llc
- Reason for Recall:
- Incorrect drill protocol in the Instructions for Use (IFU)
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Instructions for Use (IFU) and Manual NobPrl CC Guided Surgery Kit Instructions for Use (IFU): "NobelGuide for NobelParallel Conical Connection", id # 2002 and Manual: Guided Surgery with NobelParallel Conical Connection", id # 79423. Article # 38072, Nobel Parallel CC Guided Surgery Kit. The Guided Surgery Kit is used as part of the installation process of NobelParallel CC Implants. The purpose of the IFU is to instruct the user how to use the products safely. The purpose of the manual is to guide the user through the usage of NobelGuide for NobelParallel CC.
Product Codes/Lot Numbers:
Article 38072-all lots
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2375-2016
Related Recalls
NobelReplace Conical Connection RP 4.3x11.5 mm REF 36707 STERILE R
Nobel Biocare Usa
Due to 16 complaints related to drivers not engaging in implants. According to the firm's investigations, the incorrect offset gauge was potentially used during production of the implants. As a result, this could potentially lead to aborted or a delay in surgical procedure.
NobelActive Internal NP 3.5x15mm Endosseus dental implant, REF: 34128, Rx Only, Sterile R, CE, UDI: (01)07332747010619
Nobel Biocare Usa
The top label for dental implant may incorrectly indicate the implant length. Specifically, the label may contain 11.5mm for the length, which is shorter than the actual length of the implant (15mm). An incorrect sized implant may cause patient discomfort, permanent nerve damage and may require an additional surgery.
Packaged dental implant healing abutments were found to contain the wrong size device.