Volker 2080, 2082, S 280 and S 282 series beds with lateral release button. A/C Powered Adjustable Hospital Bed Product Usage: Volker model 2080/3010/3080 nursing beds are intended for the care of residents in rooms in nursing homes, in residential homes for the elderly and in suitable rooms in private residences. They may not, however, be used for transportation purposes. Any deviation from this use is expressly excluded from possible liability claims.

Class I - Dangerous
🏥 Medical Devices Recalled: June 14, 2016 Hill-Rom Infusion Pumps Nationwide

What Should You Do?

  1. Check if you have this product:
    All beds (serial numbers) in the 2080, 2082, S 280 and S 282 series are affected by this field action except for those without siderails. The models are as follows: 2080, 2082, 2080 C, 2080 E, 2080 FS, 2080 K, 2080 LP, 2080 MA, 2080 MB, 2080 S, 2080 SL, 2080 SONDER, 2082 SONDER, S280, S280 SONDER, S282, S282 SONDER.
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Hill-Rom, Inc.
Reason for Recall:
The current construction of the side rail latch may not be able to support the full patient weight and may give way allowing the side rail to abruptly lower and the patient to fall, potentially resulting in harm. To date there has been one report of a related injury.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Volker 2080, 2082, S 280 and S 282 series beds with lateral release button. A/C Powered Adjustable Hospital Bed Product Usage: Volker model 2080/3010/3080 nursing beds are intended for the care of residents in rooms in nursing homes, in residential homes for the elderly and in suitable rooms in private residences. They may not, however, be used for transportation purposes. Any deviation from this use is expressly excluded from possible liability claims.

Product Codes/Lot Numbers:

All beds (serial numbers) in the 2080, 2082, S 280 and S 282 series are affected by this field action except for those without siderails. The models are as follows: 2080, 2082, 2080 C, 2080 E, 2080 FS, 2080 K, 2080 LP, 2080 MA, 2080 MB, 2080 S, 2080 SL, 2080 SONDER, 2082 SONDER, S280, S280 SONDER, S282, S282 SONDER.

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-2376-2016

Related Recalls

Traverse rail carriage delivered with non-conforming screws are too short to ensure appropriate installation. There is potential that the screws are unable to secure the traverse rail appropriately and could loosen over time causing the rail system to fall, potentially causing injury to caregivers and/or patients

Jan 14, 2022 Other Medical Devices Nationwide View Details →