Spectrum IQ Infusion Pump, Product Code 3570009

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    UDI-DI: 00085412610900; Serial Numbers: 3815255 3815261 3815265 3815266 3815267 3815271
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Baxter Healthcare Corporation
Reason for Recall:
Baxter has identified that certain pumps have potentially been released with the grease applied to the cam and motor gears that will break down quickly. The breakdown of the grease will lead to the device having insufficient or ineffective grease applied to the cam, which could lead to premature wear of the mechanism assembly, resulting in excessive therapy.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

Spectrum IQ Infusion Pump, Product Code 3570009

Product Codes/Lot Numbers:

UDI-DI: 00085412610900; Serial Numbers: 3815255 3815261 3815265 3815266 3815267 3815271

Distribution:

Distributed in: US, AZ, FL

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-2415-2025

Related Recalls

Affected oral/axillary probes were inadvertently programmed with the rectal probe configuration. Affected probes will potentially display a lower temperature than the true value, which may lead to a delay in managing fever-related symptoms or infection.

Sep 17, 2025 Diagnostic Equipment Nationwide View Details →