AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) LAP CHOLE PACK, Model/Item Number CILC69L; 2) LAP CHOLE PACK, Model/Item Number HNLC80J; 3) LAP CHOLE PACK, Model/Item Number JKLC55H; 4) LAPAROSCOPY PACK, Model/Item Number LLLA21K; 5) GENERAL LAPAROSCOPY MPH, Model/Item Number MPGL15G;

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    1) Model/Item Number CILC69L: UDI/DI 00191072235503, Kit Lot Numbers: 42-7937111, 42-8049311; 2) Model/Item Number HNLC80J: UDI/DI 00191072193650, Kit Lot Numbers: 42-8044511, 42-8118911, 42-7977011, 42-7803211; 3) Model/Item Number JKLC55H: UDI/DI 00191072235138, Kit Lot Numbers: 42-7814311; 4) Model/Item Number LLLA21K: UDI/DI 00191072233486, Kit Lot Numbers: 70-052582, 70-053048, 70-051564, 70-051373, 70-051163, 70-053446, 70-053776; 5) Model/Item Number MPGL15G: UDI/DI 00191072223197, Kit Lot Numbers: 5436916, 5436615
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
American Contract Systems Inc
Reason for Recall:
During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) LAP CHOLE PACK, Model/Item Number CILC69L; 2) LAP CHOLE PACK, Model/Item Number HNLC80J; 3) LAP CHOLE PACK, Model/Item Number JKLC55H; 4) LAPAROSCOPY PACK, Model/Item Number LLLA21K; 5) GENERAL LAPAROSCOPY MPH, Model/Item Number MPGL15G;

Product Codes/Lot Numbers:

1) Model/Item Number CILC69L: UDI/DI 00191072235503, Kit Lot Numbers: 42-7937111, 42-8049311; 2) Model/Item Number HNLC80J: UDI/DI 00191072193650, Kit Lot Numbers: 42-8044511, 42-8118911, 42-7977011, 42-7803211; 3) Model/Item Number JKLC55H: UDI/DI 00191072235138, Kit Lot Numbers: 42-7814311; 4) Model/Item Number LLLA21K: UDI/DI 00191072233486, Kit Lot Numbers: 70-052582, 70-053048, 70-051564, 70-051373, 70-051163, 70-053446, 70-053776; 5) Model/Item Number MPGL15G: UDI/DI 00191072223197, Kit Lot Numbers: 5436916, 5436615

Distribution:

Distributed in: US, IL, LA

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-2465-2025

Related Recalls

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Sep 4, 2025 Other Medical Devices Nationwide View Details →

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Sep 4, 2025 Infusion Pumps Nationwide View Details →

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Sep 4, 2025 Other Medical Devices Nationwide View Details →