BIOFIRE FILMARRAY Gastrointestinal (GI) Panel REF RFIT-ASY-0116 (30 test kit) BIOFIRE FILMARRAY Gastrointestinal (GI) Panel REF RFIT-ASY-0104 (6 kit test) The BIOFIRE FILMARRAY Gastrointestinal (GI) Panel is a qualitative multiplexed nucleic acid-based in vitro diagnostic test intended for use with BIOFIRE¿ FILMARRAY¿ Systems. The BIOFIRE GI Panel is capable of the simultaneous detection and identification of nucleic acids from multiple bacteria, viruses, and parasites directly from stool samples in Cary Blair transport media obtained from individuals with signs and/or symptoms of gastrointestinal infection.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Part Number: RFIT-ASY-0116 (30 test kit) UDI-DI code: 00815381020109 0081538102011: All unexpired products are potentially impacted if used in conjunction with the Remel Cary Blair transport medium Part Number: RFIT-ASY-0104 (6 test kit) UDI-DI code: 0081538102011 All unexpired products are potentially impacted if used in conjunction with the Remel Cary Blair transport medium Remel Cary-Blair Transport Medium lot numbers: 743930 769240 712258 782425 732357 742049
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- BioFire Diagnostics, LLC
- Reason for Recall:
- Due to increase of false positive Vibrio/Vibrio Cholerae results when using a certain transport medium.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
BIOFIRE FILMARRAY Gastrointestinal (GI) Panel REF RFIT-ASY-0116 (30 test kit) BIOFIRE FILMARRAY Gastrointestinal (GI) Panel REF RFIT-ASY-0104 (6 kit test) The BIOFIRE FILMARRAY Gastrointestinal (GI) Panel is a qualitative multiplexed nucleic acid-based in vitro diagnostic test intended for use with BIOFIRE¿ FILMARRAY¿ Systems. The BIOFIRE GI Panel is capable of the simultaneous detection and identification of nucleic acids from multiple bacteria, viruses, and parasites directly from stool samples in Cary Blair transport media obtained from individuals with signs and/or symptoms of gastrointestinal infection.
Product Codes/Lot Numbers:
Part Number: RFIT-ASY-0116 (30 test kit) UDI-DI code: 00815381020109 0081538102011: All unexpired products are potentially impacted if used in conjunction with the Remel Cary Blair transport medium Part Number: RFIT-ASY-0104 (6 test kit) UDI-DI code: 0081538102011 All unexpired products are potentially impacted if used in conjunction with the Remel Cary Blair transport medium Remel Cary-Blair Transport Medium lot numbers: 743930 769240 712258 782425 732357 742049
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2514-2024
Related Recalls
BioFire Respiratory Panel 2.1, IVD, REF: 423742 (30 test kit)
BioFire Diagnostics
Due to unintended movement of pouch chemistry, nucleic acid test may result in elevated risk of false negative results and risk of control failures.
BioFire Respiratory Panel 2.1 plus Kit, REF: 423740, For Filmarray systems
BioFire Diagnostics
Due to manufacturing error, respiratory panel may result in elevated control failures and/or false negative results.
FilmArra Pneumonia Panel plus (Pneumoplus), REF: RFIT-LBL-0374
BioFire Diagnostics
Increased risk of control failures and false negative test results with multiplexed nucleic acid test.