Hamilton C1/T1/MR1, REF: MSP161502/12, Control Board Spare part for ventilator
Class I - DangerousWhat Should You Do?
- Check if you have this product: UDI: 07630002805247 / Affected Serial Numbers: 4001901, 4002139, 4003445, 4003567, 4003576, 4005595
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Hamilton Medical, Inc.
- Reason for Recall:
- Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Hamilton C1/T1/MR1, REF: MSP161502/12, Control Board Spare part for ventilator
Product Codes/Lot Numbers:
UDI: 07630002805247 / Affected Serial Numbers: 4001901, 4002139, 4003445, 4003567, 4003576, 4005595
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2529-2023
Related Recalls
Hamilton-T1, REF: 161006, Intensive Care and Transport Ventilator
Hamilton Medical
Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation.
Hamilton-MR1, REF: 161010, Intensive Care Ventilator
Hamilton Medical
Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation.
Hamilton-C1, REF: 161001, Portable Intensive Care Ventilator
Hamilton Medical
Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation.