OEC UroView 2800, fluoroscopic x-ray system
Class I - DangerousWhat Should You Do?
- Check if you have this product: monitor part numbers: 5448359-01, 5459219-01, or 5475662-03:89-3853, 82-1593, 8S-1935, 89-3356, 89-1323, 8S-2885, 82-0700, 8S-1708, 82-0420, 82-1998, 8S-2027, 82-7101-MH, 8S-2670, P6-0375-L, 8S-1614, 89-0846, 82-0576, 82-3256, 8S-2202, 89-0711,89-1927, 89-1453, 82-1504, 82-0794, 82-3494, 8S-3125, 8S-1845-RC, 89-3725, 8S-1592, 89-3129,82-1182, 82-0986, 8S-0199, 8S-3613, 82-3748, 82-2335, 8S-2889, 82-0596, 89-3020, 82-1515, 8S-0645, 8S-3614, 82-3925, 89-0416, 89-1363, 82-3258, 8S-0347
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- GE OEC Medical Systems, Inc
- Reason for Recall:
- A firmware issue in the supplied LCD workstation monitor(s). Under some circumstances, the monitor(s) may appear dark with no video displayed.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
OEC UroView 2800, fluoroscopic x-ray system
Product Codes/Lot Numbers:
monitor part numbers: 5448359-01, 5459219-01, or 5475662-03:89-3853, 82-1593, 8S-1935, 89-3356, 89-1323, 8S-2885, 82-0700, 8S-1708, 82-0420, 82-1998, 8S-2027, 82-7101-MH, 8S-2670, P6-0375-L, 8S-1614, 89-0846, 82-0576, 82-3256, 8S-2202, 89-0711,89-1927, 89-1453, 82-1504, 82-0794, 82-3494, 8S-3125, 8S-1845-RC, 89-3725, 8S-1592, 89-3129,82-1182, 82-0986, 8S-0199, 8S-3613, 82-3748, 82-2335, 8S-2889, 82-0596, 89-3020, 82-1515, 8S-0645, 8S-3614, 82-3925, 89-0416, 89-1363, 82-3258, 8S-0347
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2628-2016
Related Recalls
Due to manufacturing issue of insufficient sealing of X-ray tubes and can result in a potential oil leak and loss of fluoroscopic imaging capability.
Due to manufacturing issue of insufficient sealing of X-ray tubes and can result in a potential oil leak and loss of fluoroscopic imaging capability.
Due to manufacturing issue of insufficient sealing of X-ray tubes and can result in a potential oil leak and loss of fluoroscopic imaging capability.