Brilliance CT 16 Power, Model number 728240 Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes.
Class I - DangerousWhat Should You Do?
- Check if you have this product: 728240 -- Brilliance CT 16 Power 3100 5043 5047 5280 5345 5390 5413 5430 5495 5499 5512 6007 6009 6013 6032 6041 6045 6049 6051 6058 6064 6066 6070 6071 6077 6080 6081 6084 6089 6095 6100 6106 6109 6113 6115 6116 6123 6124 6131 6132 6134 6141 6144 6155 6161 6162 5476 6098 6120 6067
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Philips Medical Systems (Cleveland) Inc
- Reason for Recall:
- After upgrading to 3.6.7 software version via FCO72800643, during reconstruction of gated helical scans, the planned anatomy can be cut off on the CT images. This issue can occur on gated CT helical reconstruction that is not planned at iso-center (0,0). If the region of interest is not visualized in the images, a CT rescan may be performed.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Brilliance CT 16 Power, Model number 728240 Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes.
Product Codes/Lot Numbers:
728240 -- Brilliance CT 16 Power 3100 5043 5047 5280 5345 5390 5413 5430 5495 5499 5512 6007 6009 6013 6032 6041 6045 6049 6051 6058 6064 6066 6070 6071 6077 6080 6081 6084 6089 6095 6100 6106 6109 6113 6115 6116 6123 6124 6131 6132 6134 6141 6144 6155 6161 6162 5476 6098 6120 6067
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2656-2016
Related Recalls
Due to a software issue, there is a potential image error of the Region of Interest for expansion/contraction for HFP (Head First Prone), FFS (Feet First Supine) and FFP (Feet First Prone) orientations.
Pinnacle 3 with TumorLOC, (870258) Radiation Therapy Planning System Software
Philips Medical Systems (Cleveland)
Due to software issue, Radiation Therapy Planning system may provide incorrect dataset calculations when performing the "Stopping Power Ratio" (SPR) ,
Pinnacle3 Radiation Therapy Planning System, Model numbers 870231 and 870237.
Philips Medical Systems (Cleveland)
When computing a radiation dose in the system, the exported dose information is incorrect when there is more than one beam attached to the prescription and certain options are selected on the Edit Prescription screen.