MAGNETOM Vida Fit. Model Number: 11410481.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Model Number: 11410481. UDI Numbers: (01)04056869245188(21)202129, (01)04056869245188(21)202025, (01)04056869245188(21)202139, (01)04056869245188(21)202016, (01)04056869245188(21)202051, (01)04056869245188(21)202055, (01)04056869245188(21)202114, (01)04056869245188(21)202119, (01)04056869245188(21)202013, (01)04056869245188(21)202050, (01)04056869245188(21)202102, (01)04056869245188(21)202160, (01)04056869245188(21)202127, (01)04056869245188(21)202071, (01)04056869245188(21)202078, (01)04056869245188(21)202123, (01)04056869245188(21)202082, (01)04056869245188(21)202107, (01)04056869245188(21)202031, (01)04056869245188(21)202110, (01)04056869245188(21)202108, (01)04056869245188(21)202101, (01)04056869245188(21)202093, (01)04056869245188(21)202104, (01)04056869245188(21)202028, (01)04056869245188(21)202026, (01)04056869245188(21)202076, (01)04056869245188(21)202142, (01)04056869245188(21)202125, (01)04056869245188(21)202155, (01)04056869245188(21)202084, (01)04056869245188(21)202151, (01)04056869245188(21)202068, (01)04056869245188(21)202135, (01)04056869245188(21)202121, (01)04056869245188(21)202006, (01)04056869245188(21)202010, (01)04056869245188(21)202011, (01)04056869245188(21)202146, (01)04056869245188(21)202145. Serial Numbers: 202129, 202025, 202139, 202016, 202051, 202055, 202114, 202119, 202013, 202050, 202102, 202160, 202127, 202071, 202078, 202123, 202082, 202107, 202031, 202110, 202108, 202101, 202093, 202104, 202028, 202026, 202076, 202142, 202125, 202155, 202084, 202151, 202068, 202135, 202121, 202006, 202010, 202011, 202146, 202145.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Siemens Medical Solutions USA, Inc
- Reason for Recall:
- There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
MAGNETOM Vida Fit. Model Number: 11410481.
Product Codes/Lot Numbers:
Model Number: 11410481. UDI Numbers: (01)04056869245188(21)202129, (01)04056869245188(21)202025, (01)04056869245188(21)202139, (01)04056869245188(21)202016, (01)04056869245188(21)202051, (01)04056869245188(21)202055, (01)04056869245188(21)202114, (01)04056869245188(21)202119, (01)04056869245188(21)202013, (01)04056869245188(21)202050, (01)04056869245188(21)202102, (01)04056869245188(21)202160, (01)04056869245188(21)202127, (01)04056869245188(21)202071, (01)04056869245188(21)202078, (01)04056869245188(21)202123, (01)04056869245188(21)202082, (01)04056869245188(21)202107, (01)04056869245188(21)202031, (01)04056869245188(21)202110, (01)04056869245188(21)202108, (01)04056869245188(21)202101, (01)04056869245188(21)202093, (01)04056869245188(21)202104, (01)04056869245188(21)202028, (01)04056869245188(21)202026, (01)04056869245188(21)202076, (01)04056869245188(21)202142, (01)04056869245188(21)202125, (01)04056869245188(21)202155, (01)04056869245188(21)202084, (01)04056869245188(21)202151, (01)04056869245188(21)202068, (01)04056869245188(21)202135, (01)04056869245188(21)202121, (01)04056869245188(21)202006, (01)04056869245188(21)202010, (01)04056869245188(21)202011, (01)04056869245188(21)202146, (01)04056869245188(21)202145. Serial Numbers: 202129, 202025, 202139, 202016, 202051, 202055, 202114, 202119, 202013, 202050, 202102, 202160, 202127, 202071, 202078, 202123, 202082, 202107, 202031, 202110, 202108, 202101, 202093, 202104, 202028, 202026, 202076, 202142, 202125, 202155, 202084, 202151, 202068, 202135, 202121, 202006, 202010, 202011, 202146, 202145.
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2668-2025
Related Recalls
MAMMOMAT Inspiration;
Siemens Medical Solutions USA
There were instances where the operator table was sold together with a bus-installation kit. The operator table is not intended and not released for bus installations as it was designed for stationary operation only.
MAMMOMAT Fusion;
Siemens Medical Solutions USA
There were instances where the operator table was sold together with a bus-installation kit. The operator table is not intended and not released for bus installations as it was designed for stationary operation only.
MAMMOMAT Revelation;
Siemens Medical Solutions USA
There were instances where the operator table was sold together with a bus-installation kit. The operator table is not intended and not released for bus installations as it was designed for stationary operation only.