NOMAD Pro2 Packaged X-Ray System, Part Number 08500021; NOMAD Pro2 Packaged X-Ray System Black, Part Number 08500022; NOMAD Pro2 Vet Packaged X-Ray System, Part Number 08500023; NOMAD Pro2 Vet Packaged X-Ray System Black, Part Number 08500024; NOMAD Pro2 Packaged X-Ray System, GBL, Part Number 08500025 and Part Number 08500026; NOMAD Pro2 Packaged X-Ray System, GLB, Vet, Part Number 08500027 and Part Number 08500028; NOMAD Pro2 Package X-Ray System, GLB, Black, Part Number 08500029
Class I - DangerousWhat Should You Do?
- Check if you have this product: Part Number 08500021 - Serial Number Range XD-100921-05 - XD-150109-11, 10000 - 20811, XD-140929-39 - XD-150123-20; Part Number 08500022 - Serial Number Range XE-130528-03 - XE-150123-38, 10000 - 20541, XE-140908-32 - XE-150123-40; Part Number 08500023 - Serial Number Range XF-121008-05 - XF-150205-03, 10003 - 20032, XE-150205-01; Part Number 08500024 - Serial Number Range XG-130520-01 - XG-141210-05, 1000 - 20019; Part Number 08500025 - Serial Number Range XI-140303-02 - XI-150210-9, XI-150202-04 - XI-150210-19; Part Number 08500026 - Serial Number Range 10039 - 20041; Part Number 08500027 - Serial Number Range 1000 - 20006, Part Number 08500028 - Serial Number Range 10000 - 20030; Part Number 08500029 - Serial Number Range 10000.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Aribex Inc
- Reason for Recall:
- Firm received customer complaints where end users reported that Nomad Pro and Pro 2 handsets had appeared to "overheat", "melt", "smoke" and/or "ignite." Investigation found that the internal lithium ion batteries had entered into a thermal runaway condition as a result of a short circuit on the battery pack PCB.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
NOMAD Pro2 Packaged X-Ray System, Part Number 08500021; NOMAD Pro2 Packaged X-Ray System Black, Part Number 08500022; NOMAD Pro2 Vet Packaged X-Ray System, Part Number 08500023; NOMAD Pro2 Vet Packaged X-Ray System Black, Part Number 08500024; NOMAD Pro2 Packaged X-Ray System, GBL, Part Number 08500025 and Part Number 08500026; NOMAD Pro2 Packaged X-Ray System, GLB, Vet, Part Number 08500027 and Part Number 08500028; NOMAD Pro2 Package X-Ray System, GLB, Black, Part Number 08500029
Product Codes/Lot Numbers:
Part Number 08500021 - Serial Number Range XD-100921-05 - XD-150109-11, 10000 - 20811, XD-140929-39 - XD-150123-20; Part Number 08500022 - Serial Number Range XE-130528-03 - XE-150123-38, 10000 - 20541, XE-140908-32 - XE-150123-40; Part Number 08500023 - Serial Number Range XF-121008-05 - XF-150205-03, 10003 - 20032, XE-150205-01; Part Number 08500024 - Serial Number Range XG-130520-01 - XG-141210-05, 1000 - 20019; Part Number 08500025 - Serial Number Range XI-140303-02 - XI-150210-9, XI-150202-04 - XI-150210-19; Part Number 08500026 - Serial Number Range 10039 - 20041; Part Number 08500027 - Serial Number Range 1000 - 20006, Part Number 08500028 - Serial Number Range 10000 - 20030; Part Number 08500029 - Serial Number Range 10000.
Distribution:
Nationwide distribution
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2717-2016
Related Recalls
Firm received customer complaints where end users reported that Nomad Pro and Pro 2 handsets had appeared to "overheat", "melt", "smoke" and/or "ignite." Investigation found that the internal lithium ion batteries had entered into a thermal runaway condition as a result of a short circuit on the battery pack PCB.