ThruPort Knot Pusher. This device is sold individually (Model KP1) and as part of the ThruPort Valve Placement Pack (Model VPP). packaged sterile and non-pyrogenic in a sealed, peel-type pouch. The Knot Pusher is a sterile disposable plastic rod used to advance extracorporeal suture knots during minimally invasive cardiac surgery.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Model: KP1 - Lot No: 59964698, 59967596, 59984375, 60016312, 60045801, 60045802, 60059982, 60066591, 60089865 Model VPP Lot No: 60022521, 60022522, 60064112
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Edwards Lifesciences, LLC
Reason for Recall:
The configuration of the slot at the tip of the knot pusher may inhibit intended ease of use. Fraying or splitting of the suture could occur, which could compromise the integrity of the suture knot.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

ThruPort Knot Pusher. This device is sold individually (Model KP1) and as part of the ThruPort Valve Placement Pack (Model VPP). packaged sterile and non-pyrogenic in a sealed, peel-type pouch. The Knot Pusher is a sterile disposable plastic rod used to advance extracorporeal suture knots during minimally invasive cardiac surgery.

Product Codes/Lot Numbers:

Model: KP1 - Lot No: 59964698, 59967596, 59984375, 60016312, 60045801, 60045802, 60059982, 60066591, 60089865 Model VPP Lot No: 60022521, 60022522, 60064112

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-2780-2015

Related Recalls

due to complaints received for 3mm to 4mm section wire exposed/protruding from the outside of the cannula body.

May 14, 2025 Other Medical Devices Nationwide View Details →

due to complaints received for 3mm to 4mm section wire exposed/protruding from the outside of the cannula body.

May 14, 2025 Other Medical Devices Nationwide View Details →

due to complaints received for 3mm to 4mm section wire exposed/protruding from the outside of the cannula body.

May 14, 2025 Other Medical Devices Nationwide View Details →