Zimmer Guide Wires for orthopedic procedures.
Class I - DangerousWhat Should You Do?
- Check if you have this product: 00-2228-024-00 2.4MM Diameter Bullet Tip Guide Wire; 00-2255-025-00 Humeral Smooth Guide Wire 2.4mm 70cm Length; 00-2255-026-00 Humeral Bullet Tip Guide Wire 2.4mm Diameter 70cm Length; 47-2237-033-00 Smooth Guide Wire 2.4mm Diameter 100cm Length; 47-2237-037-00 3.0MM Diameter Smooth Guide Wire 100cm Length; 47-2237-038-00 Bullet Tip Guide Wire 3.0mm Diameter 100cm Length; 47-2255-008-00 Ball Tip Guide Wire 2.4mm Diameter 70cm Length; 47-2255-008-01 Ball Tip Guide Wire 3.0mm Diameter 100cm Length All non-expired, distributed lots with a 10 year expiration date on the label.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
ā ļø Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Zimmer, Inc.
- Reason for Recall:
- Product labeling of these sterile guide wires (external carton label and patient record label) states a 10-year expiration date. Package testing supports a 5-year expiration date for these same guide wires.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Zimmer Guide Wires for orthopedic procedures.
Product Codes/Lot Numbers:
00-2228-024-00 2.4MM Diameter Bullet Tip Guide Wire; 00-2255-025-00 Humeral Smooth Guide Wire 2.4mm 70cm Length; 00-2255-026-00 Humeral Bullet Tip Guide Wire 2.4mm Diameter 70cm Length; 47-2237-033-00 Smooth Guide Wire 2.4mm Diameter 100cm Length; 47-2237-037-00 3.0MM Diameter Smooth Guide Wire 100cm Length; 47-2237-038-00 Bullet Tip Guide Wire 3.0mm Diameter 100cm Length; 47-2255-008-00 Ball Tip Guide Wire 2.4mm Diameter 70cm Length; 47-2255-008-01 Ball Tip Guide Wire 3.0mm Diameter 100cm Length All non-expired, distributed lots with a 10 year expiration date on the label.
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2820-2015
Related Recalls
Zimmer Biomet is conducting a medical device recall for two (2) lots of the Dual Mobility Vivacit-EĀæ Bearing due to a commingle. The outer package is labeled as a Size F 44 mm, however, the implant inside the package is a Size G 46 mm, and vice versa.
Zimmer Biomet is conducting a medical device recall for two (2) lots of the Dual Mobility Vivacit-EĀæ Bearing due to a commingle. The outer package is labeled as a Size F 44 mm, however, the implant inside the package is a Size G 46 mm, and vice versa.