The da Vinci Xi Surgical System (IS4000) Model Name: IS4000, Endoscope Controller. Model Number: 372601-16.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    UDI Number: 00886874114193. Serial Number: SK1703.
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Intuitive Surgical, Inc.
Reason for Recall:
An Endoscope Controller has been improperly calibrated during manufacturing. This improper calibration may result in the Endoscope Controller showing surgical images that are dimmer, have color imbalance or have reduced fluorescence and grayscale when using Firefly. There is also potential for increased fogging of the endoscope.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

The da Vinci Xi Surgical System (IS4000) Model Name: IS4000, Endoscope Controller. Model Number: 372601-16.

Product Codes/Lot Numbers:

UDI Number: 00886874114193. Serial Number: SK1703.