- Company:
- Abiomed, Inc.
- Reason for Recall:
- A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps during coronary interventions. Currently there is no Abiomed labeling (IFU etc.) content related to this potential interaction, so Abiomed will update the IFU to include information on this potential device interaction.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Impella CP Pump Set, Japan; Product Code: 0048-0034-JP;
Product Codes/Lot Numbers:
Product Code: 0048-0034-JP; UDI-DI: 00813502011463; Serial Numbers: 324303 323153 328083 328085; Batch Numbers: 2022079852 2022079846 2022083745 2022083747
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2956-2024
Related Recalls
Potential for Automated Impella Controller (AIC) purge pressure issues due to purge retainer failures.
Potential cybersecurity vulnerabilities related to the operating system in the Automated Impella Controller (AIC).
Specific Automated Impella Controllers (AICs) that have a Pump Driver Circuit Assembly do not meet current specifications. This may lead to decreased pump performance or pump stop and trigger an Impella Failure or Impella Stopped Controller Failure alarm.