400mL, Premium, Urine Meter, 14 FR, Silicone Temperature Sensing Foley Traywith Cath-Secure Plus, Sterile, Item Code P4P14TSD
Class I - DangerousWhat Should You Do?
- Check if you have this product: 1733152764 1806451564
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- COVIDIEN LLC
- Reason for Recall:
- The temperature sensor catheters may be defective in that they will show a lower body temperature measurement than the actual temperature of the patient.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
400mL, Premium, Urine Meter, 14 FR, Silicone Temperature Sensing Foley Traywith Cath-Secure Plus, Sterile, Item Code P4P14TSD
Product Codes/Lot Numbers:
1733152764 1806451564
Distribution:
Nationwide distribution
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-3205-2018
Related Recalls
Temperature probe devices lack FDA clearance.
Due to customer complaint regarding incorrect display box labeling.