Incisive CT; Software Version Number: 4.5, 5.0, 5.1;
Philips North America
If the screws of tube heat exchanger on the rotating scanner are not tightened well after replacing. This component may become detached and make contact with other components located within the Incisive CT systems during rotation. Other components could be damaged due to the contact with the tube heat exchanger.