Firm distributes kits which include the BD MaxZeroNeedleess Connector, which was recalled by the manufacturer due to the potential for droplets of fluid separating from the sealed surface of the connector following disconnection.
Flow cytometers, when used with specific tubes may experience excessive electronic abort counts (electronic abort count of beads exceeds 1.0% of the total processed events), which may affect the ratio of cell population events to bead events, potentially resulting in falsely high absolute counts. This may affect the following: initiation of prophylactic therapy for Opportunistic Infections (HIV patients), decisions regarding care/management (PID/immune suppressed patients), laboratory developed tests/experiments, specimen engraftment potency, and use of leucoreduced blood products for transfusion.
Firm distributes kits which include the BD MaxZeroNeedleess Connector, which was recalled by the manufacturer due to the potential for droplets of fluid separating from the sealed surface of the connector following disconnection.
The action is being initiated because the cooling indicator board was improperly installed. This was first noticed by a service employee during assembly. If for some reason the cooling of the reagent rotor fails, the reagent cooling indicator light will not illuminate and the acoustic beeper will not go off. As a consequence the operator will not be warned and the reagent storage conditions may fall out of specification.
Firm distributes kits which include the BD MaxZeroNeedleess Connector, which was recalled by the manufacturer due to the potential for droplets of fluid separating from the sealed surface of the connector following disconnection.
Firm distributes kits which include the BD MaxZeroNeedleess Connector, which was recalled by the manufacturer due to the potential for droplets of fluid separating from the sealed surface of the connector following disconnection.
A mismatch between the angle of the hole in the base and the angle of the bell, which may cause the clamp to remove foreskin without the use of a scalpel.
Firm distributes kits which include the BD MaxZeroNeedleess Connector, which was recalled by the manufacturer due to the potential for droplets of fluid separating from the sealed surface of the connector following disconnection.
Firm distributes kits which include the BD MaxZeroNeedleess Connector, which was recalled by the manufacturer due to the potential for droplets of fluid separating from the sealed surface of the connector following disconnection.
AssayTips part of the AssayTip/AssayCup Tray Tips used on the cobas e 801 module- may have an Abnormal Internal Structure and lead to Incorrect patient results
Firm distributes kits which include the BD MaxZeroNeedleess Connector, which was recalled by the manufacturer due to the potential for droplets of fluid separating from the sealed surface of the connector following disconnection.
Firm distributes kits which include the BD MaxZeroNeedleess Connector, which was recalled by the manufacturer due to the potential for droplets of fluid separating from the sealed surface of the connector following disconnection.
Firm distributes kits which include the BD MaxZeroNeedleess Connector, which was recalled by the manufacturer due to the potential for droplets of fluid separating from the sealed surface of the connector following disconnection.
The outer seam of the coil, closest to the system cable, may separate and expose the porous material internal to the coil. This could result in user and/or patient contact with infectious agents since the internal porous material cannot be disinfected.