During manufacturing, the screw stop ledge was not made to specification and location within the abutment, resulting in the screw head seating deeper in the abutment with no positive stop.
All Recalls
95,123 total recalls in our database
Reports when using Rheumatoid Factor test kits, control and negative samples may be elevated following a sample containing high levels of IgG . The negative control may report out of range; a negative sample may recover as a low positive which could contribute to an incorrect or premature diagnosis of rheumatoid.
A voluntary recall of a lot (67924) of Independent Care System due to the potential for damaged Sodium Bicarbonate Mouthpaste (cosmetic) tubes.
When unscrewing the green cap from the pen, a visible piece of green plastic was seen on the outer barrel of the pen.
Devices were distributed despite the Helium-fine leak test results not being within the cochlear implant standards specification.
The cold restart of Azurion R1.2 systems may take up to 7 minutes if the system is connected to the mains power supply for more than 50 days, may result in a delay of treatment.
iCT, Model # 728306, computed tomography x-ray system
Philips Medical Systems (Cleveland)
In certain instances when performing a cardiac step and shoot acquisition with phase tolerance selected, images may be labeled as if phase tolerance had been applied when phase tolerance had not been applied
Potentially Biased Results using VITROS Chemistry Products GLU Slides
VITROS 250 Chemistry System, clinical chemistry analyzer
Ortho Clinical Diagnostics
Potential Unsuccessful Calibration Due to Software Anomaly in VITROS System
Electrical failures were identified in cardioverter defibrillators (ICDs) due to damaged aluminum wires. ICDs may contain electrical wire connections which may not be completely insulated. The potential patient impact could be the inability to deliver high voltage therapy. There is no available option to verify the vulnerability status for implanted devices.
YXLON Cougar Cabinet X-Ray System Product Usage: The products are generally used for non-destructive testing and quality control
Comet Technologies Usa
Failure to meet the requirement of the performance standard for cabinet x-ray systems (21 CFR 1020.40) for a door interlock that causes physical disconnection of the power supply using only the movement of the door.
There is an increased chance (approx. 5%) of no electrolyte or a reduced amount of electrolyte being applied to the sensor during the membrane change procedure.
ELLIPSE DR, Tiered-therapy cardioverter/defibrillator, REF: CD2411-36C, UDI: 05414734507585
St Jude Medical
Electrical failures were identified in cardioverter defibrillators (ICDs) due to damaged aluminum wires. ICDs may contain electrical wire connections which may not be completely insulated. The potential patient impact could be the inability to deliver high voltage therapy. There is no available option to verify the vulnerability status for implanted devices.
YXLON Cheetah cabinet X-Ray System Product Usage: The products are generally used for non-destructive testing and quality control
Comet Technologies Usa
Failure to meet the requirement of the performance standard for cabinet x-ray systems (21 CFR 1020.40) for a door interlock that causes physical disconnection of the power supply using only the movement of the door.
14F x 24CM SLX Double Lumen Full Tray Hemodialysis catheter insertion kits, Item No. MC061424 Product Usage: An aid for introducing a hemodialysis catheter
Medical Components, Inc
Tray label indicates the kits are packaged with a 15F "valved" peelable introducer, but the kits contain a 15F "non-valved" peelable introducer.
LACERATION TRAY Kit Code: SUT18370 - Product Usage: Kit is used for laceration repairs for general population.
Centurion Medical Products
Potential for sterile packaging to be compromised
SNAG FREE SUTURE TRAY Kit Code: SUT14355 - Product Usage: Kit is used for laceration repairs for general population.
Centurion Medical Products
Potential for sterile packaging to be compromised
There has been an increase in complaints related to leaks, suction issues, and failure to obtain samples while using the product.
Ingenuity CT, Model # 728326, computed tomography x-ray system
Philips Medical Systems (Cleveland)
In certain instances when performing a cardiac step and shoot acquisition with phase tolerance selected, images may be labeled as if phase tolerance had been applied when phase tolerance had not been applied
UNIVERSAL BLOCK TRAY Kit Code: MNS6645 - Product Usage: Kit is used for regional anesthesia nerve blocks for general population.
Centurion Medical Products
Potential for sterile packaging to be compromised