Labeling: Label Error on Declared Strength; Some cartons were incorrectly labeled. The blister strips inside the product carton reflect the correct strength.
All Recalls
95,114 total recalls in our database
Presence of Foreign Tablets/Capsules: A Paracetamol 500 mg tablet was found in a 1000-count bottle of Metformin HCL ER Tablets USP, 500 mg.
Labeling: Label Error on Declared Strength; Some cartons were incorrectly labeled. The blister strips inside the product carton reflect the correct strength.
Duloxetine Delayed-Release Capsules USP 20 mg, 60 count bottles, Rx only, Distributed by: Rising Pharm Holdings, Inc., East Brunswick, NJ NDC 57237-017-60
Rising Pharma Holding
CGMP Deviations; Presence of N-nitroso-duloxetine impurity above recommended interim limit.
Tec 820 ISO Anesthesia Vaporizer, Reference Number 1177-9820-ISO
GE Medical Systems
Due to a manufacturing issue, certain Tec 6 Plus and Tec 800 series vaporizers can deliver less anesthetic agent than the vaporizer dial setting.
Tec 850 SEV Anesthesia Vaporizer, Reference Number 1177-9850-SEV
GE Medical Systems
Due to a manufacturing issue, certain Tec 6 Plus and Tec 800 series vaporizers can deliver less anesthetic agent than the vaporizer dial setting.
Beckman Coulter has identified that some Access PTH reagent packs from lot 339071 may be over- or under-filled in well 0. The dispensed material in well 0 could be above or below specification, risking sealing issues, contamination, instrument errors, delays, or inaccurate results with a negative bias and increase in imprecision. The issue was confirmed on 10 April 2024 by Beckman Coulter through an internal non-conformance (NC-INT-66585). The well 0 scale on the filling line began malfunctioning on the fill immediately before PTH. The malfunction was not noticed until the majority of PTH was filled.
GE Healthcare Optima XR220amx, Mobile Digital Radiographic System, Model Numbers: 1) 5555000-12; 2) 5555000-5; 3) 5555000-6; 4) A0659PB; 5) to be provided
GE Medical Systems
GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement performed, some bolts used in the Column Safety mechanism can be missing. If these bolts are missing and a failure of the internal counterpoise cable occurs, the Column Safety mechanism will not work properly causing the arm with X-ray tube and collimator to drop downward on the column track.
GE Healthcare Optima XR200amx, Mobile Digital Radiographic System, Model Numbers: 1) 5555000-3; 2) 5555000-4: 3) A0659MB
GE Medical Systems
GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement performed, some bolts used in the Column Safety mechanism can be missing. If these bolts are missing and a failure of the internal counterpoise cable occurs, the Column Safety mechanism will not work properly causing the arm with X-ray tube and collimator to drop downward on the column track.
Ivenix Infusion System (IIS), Large Volume Pump, Model Number: LVP-0004
Fresenius Kabi USA
Potential pneumatic valve failure that may cause the LVP to stop an active infusion and/or could prevent the pump from being used.
Aerin Medical VivAer Stylus, REF CAT1785 (individual stylus), and REF CAT500 (5-pack), sterile.
Aerin Medical
Some units in a single production lot were programmed incorrectly which will result in the VivAer Stylus being incorrectly recognized as RhinAer Stylus when connected to the Aerin Console.
Three lots of the FastPacks kitted into the FastPack SHBG Complete Kits were labeled with the incorrect expiration date. The following is a listing of affected lots: Lot Number: 2408018-1, 2409015-1, and 2409015-2. The likelihood of patient harm is expected to be low as routine Quality Control should detect kits that are impacted by signal deterioration due to age.
Xhibit Telemetry Receiver, Model 96280, with Service Manual. Intended to provide monitoring system with patient data.
Spacelabs Healthcare
Telemetry receiver systems may experience a Windows proxy Issue causing the system to shut down and restart (Offline displayed), lasting approximately 1 to 1.5 minutes, which will cause temporary loss of monitoring. If a clinical event occurred during this period, staff may not detect until the system restarts; necessary clinical intervention could be delayed.
The devices were inadvertently shipped after their expiration date.
There is a potential safety risk caused by physical disassociation of the back section from the bed frame of Skytron model series 3600, 6500, 6600 and 6700 operating room beds when used with the EnVisio Navigation System.
Citric Complete - Dry Citric Acid Concentrate Model Number: DCA+225-25
Nipro Renal Soultions USA
The outer case is mislabeled as 3.0 K, 2.0 Ca instead of 2.0 K, 3.0 Ca.
VACUETTE SAFELINK, REF: 450210
Greiner Bio-One GmbH
The VACUETTE SAFELINK Holder with male luer lock may leak blood due to the protective rubber sleeve bunching up and not retracting when using the device.
GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement performed, some bolts used in the Column Safety mechanism can be missing. If these bolts are missing and a failure of the internal counterpoise cable occurs, the Column Safety mechanism will not work properly causing the arm with X-ray tube and collimator to drop downward on the column track.
JMC5A Ni/TruAire-5 Oxygen Concentrator, Model: O2C5L
JIANGSU JUMAO X-CARE MEDICAL EQUIPMENT CO
Oxygen concentrator may melt or catch on fire during use.
Xhibit Telemetry Receiver with Software, Model 96280. Intended to provide monitoring system with patient data.
Spacelabs Healthcare
Telemetry receivers with affected software versions may experience unsolicited system shut down and restart (Offline displayed), lasting approximately 1 to 1.5 minutes, which will cause temporary loss of monitoring. If a clinical event occurred during this period, staff may not detect until the system restarts and necessary clinical intervention could be delayed.