Labeling: Not Elsewhere Classified: Back Label states 'Each contains: cephalexin monohydrate, USP equivalent to 5g' instead of Each Bottle contains: cephalexin monohydrate, USP equivalent to 10g'
Presence of Foreign Tablets/Capsules: Complaint received from a re-packager, American Health Packaging (AHP), where a foreign tablet was discovered in one of the bottles during packaging set up. Tablet identified as pantoprazole tablet 20mg.
Labeling: Not Elsewhere Classified: Back Label states 'Each contains: cephalexin monohydrate, USP equivalent to 5g' instead of Each Bottle contains: cephalexin monohydrate, USP equivalent to 10g'
Labeling: Not Elsewhere Classified: Front label states 100 Tablets instead of 100 ml and on the back label 'Each Tablet contains' instead of 'Each Bottle contains' No total dose per bottle listed should be 2.5 g'
Stryker Neurovascular has observed an increased frequency of PTFE coating damage occurring on the Synchro Guidewires that may be caused by the practice of backloading the guidewire through the optional introducer accessory. This issue is limited to certain lots of the Synchro Guidewire that contain an older version of the introducer accessory. Due to variation in the manufacturing process of the supplier of the introducer accessory, certain lots of introducers have sharper than intended edges that can peel off the PTFE coating when physicians use a technique known as backloading. Users with impacted product in their inventory are cautioned not to use this backloading technique.
It has been identified that testing documentation does not support that the syringes can successfully deliver accurate volumes across the full range of their claimed graduated capacity and may pose a risk to patient health.
According to 21 CFR 1040.10(f)(4), each laser system classified as a Class IIIb or IV laser product shall incorporate a key-actuated master control. The key shall be removable and the laser shall not be operable when the key is removed. However, the key of the subject laser products can be removed from key switch and the laser remains on. This is a safety concern.