CGMP Deviations: products may not conform to the labeled specifications.
All Recalls
95,114 total recalls in our database
Scorpio-Flex Ttl Kn P-S Tib Brg Insrt Asy- -Intended for knee replacement Part Number: 72-16-0910
Howmedica Osteonics
Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
PRIMARY REV.TIB.INSERT-DURACON Intended for knee replacement. Product Number: 6637-4-031
Howmedica Osteonics
Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
Specific lots of sutures were sterilized with gamma doses that exceeded the range approved. The extra gamma levels may decrease the strength of the sutures over time, which could result in harms such as wound dehiscence and hemorrhage/blood loss/ bleeding at a critical level.
PS LIPPED TIBIAL INSERT ASSY DURACON Intended for knee replacement Product Number: 6742-1-411
Howmedica Osteonics
Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
SCORPIO TOTAL KNEE MEDIALIZED DOME PATELLA-SZ #11 -Intended for knee replacement Part Number: 73-0110
Howmedica Osteonics
Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
EZDilate Wire Guided Balloon 6-7-8. indicated for endoscopic dilation of strictures in the alimentary tract in adults and adolescents (>12 years) Model: BD-410X-0855
Olympus Corporation of the Americas
Product balloon issues: inflating, deflating and/or retrieving the devices; reports of bursting and leaking and events involving foreign body in patient and prolonged procedure.
Intera 1.5T Achieva Nova. Model (REF) Numbers 781172.
Philips North America
The Quadrature Body Coil (QBC) seal adhesive may fail, creating sharp edges that may come in contact with patients. If the QBC seal becomes loose during the scanning process, the risk to the patient may include skin abrasions, bruises, lacerations, hair loss/entanglement, and tissue injury.
LEFT SMALL-PCA MOD.REV.DUR.REV.INSERT Intended for knee replacement. Product Number: 6637-0-631
Howmedica Osteonics
Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
EZDilate Wire Guided Balloon 16-17-18. Indicated for endoscopic dilation of strictures in the alimentary tract in adults and adolescents (>12 years). Model: BD-410X-1855
Olympus Corporation of the Americas
Product balloon issues: inflating, deflating and/or retrieving the devices; reports of bursting and leaking and events involving foreign body in patient and prolonged procedure.
smiths medical Medfusion Model 3500 Syringe pump
Smiths Medical ASD
Medfusion syringe pumps, with software versions before v6.0.0, may have the following issues:1. Delivery During Motor Not Running High Priority Alarm, 2. Infusion Restarted with Incorrect Parameters, 3. Screen Lock, 4. Interruption of Bolus or Loading Dose Delivery, 5. Pump Displays Incorrect Bolus/Loading Dose, 6. Loading/Bolus Dose Below the Minimum Recommended Rate, 7. Motor Rate Error, 8. Incorrect Recall Last Settings, 9. Corrupt Configuration, 10. Auto Lock, 11. Toolbox Configuration Loading Dose Time Values. Smiths Medical corrected all issues included in this notification in previous software updates and the corrections were carried forward into all subsequent software releases. Please ensure you have the most recent Medfusion software (v6.0.0) installed on your pumps.
MEDLINE BLOOD CULTURE KIT, Reorder Number DYNDH1723
MEDLINE INDUSTRIES, LP - Northfield
Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Various Adult Blood and Blood Culture Kits. This recall is due to kits built using a Bactec component lot and expiration date that were not reflected on the kit insert. There is a risk of using an expired component if the user does not check the Bactec inside the kit for its expiration date.
System may exhibit a loss of imaging functionality and data due to an issue which can cause the system to continuously restart. If the issue occurs during a procedure, there may be a delay and/or abortion of the procedure.
SCORPIO U-DOME PATELLA -Intended for knee replacement Part Number: 73-3710
Howmedica Osteonics
Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
SCORPIO TOTAL KNEE CONCENTRIC DOME PATELLA-SZ #9 -Intended for knee replacement Part Number: 73-2910
Howmedica Osteonics
Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
AtriCure has identified certain lots of tank hose assemblies where a second-tier supplier provided incorrect tank hose assemblies. These tank hose assemblies were provided with an outer sheathing that does not have required pin perforations and the inner tube of the hose is not the correct material.
GE Voluson IC9-RS intracavitary probes, Model Number J48691PJ, Diagnostic ultrasound imaging and fluid flow analysis
GE HEALTHCARE AUSTRIA GMBH &
GE HealthCare has become aware of an issue with certain IC9-RS ultrasound probes (see Affected Products List) that can result in a double image artifact creating a ghost image with realistic features. An unrecognized artifact may lead to misdiagnosis.
MEDLINE BLOOD CULTURE KIT, Reorder Number DYNDH1622C
MEDLINE INDUSTRIES, LP - Northfield
Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Various Adult Blood and Blood Culture Kits. This recall is due to kits built using a Bactec component lot and expiration date that were not reflected on the kit insert. There is a risk of using an expired component if the user does not check the Bactec inside the kit for its expiration date.
SCORPIO TOTAL KNEE CONCENTRIC DOME PATELLA-SZ #7 -Intended for knee replacement Part Number: 73-2710
Howmedica Osteonics
Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
SCORPIO U-DOME PATELLA -Intended for knee replacement Part Number: 73-3110
Howmedica Osteonics