NIBP valve assembly board subject to this recall includes enhancements that are not included in the current marketing clearance:1. Anti-motion feature 2. NIBP hose blockage identification function
Diagnostic Equipment
🏥 Medical Devices • 6,627 recalls
Mammomat Revelation with software version VC10 and Biopsy Option Intended Use: For mammography exams, screening, diagnostics, biopsies and dual energy procedures
Siemens Medical Solutions USA
An error can occur when performing a biopsy using the InSpect function. The acquisition workstation may become unresponsive to normal user interaction, can only take place in rare cases
xTAG Respiratory Viral Panel (RVP): Box Label: xTAG RVP (Respiratory Viral Panel). Catalogue Number:I019C0111 in vitro diagnostic test
Luminex Molecular Diagnostics
The Assay may give a false negative result when detecting and subtyping Influenza A (H3) and detecting RSV A in patient specimens. However, the assay will continue to detect Influenza A, Influenza B, RSV B and all other targets in accordance with each of these products performance claims.
Cerelink ICP Monitor, Model Number 826820; intended for use as an interface between compatible strain-gauge type pressure transducers and standard physiological pressure monitoring systems.
Integra LifeSciences
There is a potential for a progressive decline in observed ICP readings of the monitor during use.
Roche Diagnostics cobas Integra Albumin Gen.2 Catalog Number: 05166861190 - Product Usage: In vitro test for the quantitative determination of the albumin in human serum and plasma on Roche/Hitachi cobas c systems.
Roche Diagnostics Operations
Due to low Quality Control recovery and invalid Calibration.
Cobas c Bilirubin Total Gen.3 Catalog Number: 05795419190 - Product Usage: In vitro test for the quantitative determination of the total bilirubin in serum and plasma of adults and neonates on Roche/Hitachi cobas c systems.
Roche Diagnostics Operations
Due to low Quality Control recovery and invalid Calibration.
Siemens Atellica CH 930 Analyzer. Siemens Material Number (SMN): 11067000 Product Usage: automated, clinical chemistry analyzer to perform in vitro diagnostic tests on clinical specimens
Siemens Healthcare Diagnostics
Atellica CH 930 Analyzer - Erroneous Indices for Hemolysis and Lipemia in Software versions 1.19.2 and below
Online TDM Vancomycin Gen.3 (100 tests) Catalog # 06779336190 Analyzer/Module: cobas c 311 and cobas c 501/502
Roche Diagnostics Operations
Inaccurately Low Sample Results
BD FACSLyric Flow Cytometer, 3-Laser 12 Color Instrument, REF: 663029, when using BD Trucount Tubes
Becton, Dickinson and Company, BD Biosciences
Flow cytometers, when used with specific tubes may experience excessive electronic abort counts (electronic abort count of beads exceeds 1.0% of the total processed events), which may affect the ratio of cell population events to bead events, potentially resulting in falsely high absolute counts. This may affect the following: initiation of prophylactic therapy for Opportunistic Infections (HIV patients), decisions regarding care/management (PID/immune suppressed patients), laboratory developed tests/experiments, specimen engraftment potency, and use of leucoreduced blood products for transfusion.
BD FACSLyric Flow Cytometer, 2-Laser 4 Color Instrument, REF: 662875, when using BD Trucount Tubes
Becton, Dickinson and Company, BD Biosciences
Flow cytometers, when used with specific tubes may experience excessive electronic abort counts (electronic abort count of beads exceeds 1.0% of the total processed events), which may affect the ratio of cell population events to bead events, potentially resulting in falsely high absolute counts. This may affect the following: initiation of prophylactic therapy for Opportunistic Infections (HIV patients), decisions regarding care/management (PID/immune suppressed patients), laboratory developed tests/experiments, specimen engraftment potency, and use of leucoreduced blood products for transfusion.
Achilles Express Bone Sonometer
GE Healthcare
Certain Achilles Express and Achilles Insight systems were shipped with European CEE-7/7 Type Power Cords to countries where three protruding pin plugs are required for protective earth connection.
BD FACSLyric Flow Cytometer, 3-Laser 8 Color Instrument, REF: 662877, when using BD Trucount Tubes
Becton, Dickinson and Company, BD Biosciences
Flow cytometers, when used with specific tubes may experience excessive electronic abort counts (electronic abort count of beads exceeds 1.0% of the total processed events), which may affect the ratio of cell population events to bead events, potentially resulting in falsely high absolute counts. This may affect the following: initiation of prophylactic therapy for Opportunistic Infections (HIV patients), decisions regarding care/management (PID/immune suppressed patients), laboratory developed tests/experiments, specimen engraftment potency, and use of leucoreduced blood products for transfusion.
BD FACSLyric Flow Cytometer, 2-Laser 4 Color Instrument, REF: 651164, when using BD Trucount Tubes
Becton, Dickinson and Company, BD Biosciences
Flow cytometers, when used with specific tubes may experience excessive electronic abort counts (electronic abort count of beads exceeds 1.0% of the total processed events), which may affect the ratio of cell population events to bead events, potentially resulting in falsely high absolute counts. This may affect the following: initiation of prophylactic therapy for Opportunistic Infections (HIV patients), decisions regarding care/management (PID/immune suppressed patients), laboratory developed tests/experiments, specimen engraftment potency, and use of leucoreduced blood products for transfusion.
BD FACSLyric Flow Cytometer, 3-Laser 10 Color Instrument, REF: 662878, and 66287808 when using BD Trucount Tubes
Becton, Dickinson and Company, BD Biosciences
Flow cytometers, when used with specific tubes may experience excessive electronic abort counts (electronic abort count of beads exceeds 1.0% of the total processed events), which may affect the ratio of cell population events to bead events, potentially resulting in falsely high absolute counts. This may affect the following: initiation of prophylactic therapy for Opportunistic Infections (HIV patients), decisions regarding care/management (PID/immune suppressed patients), laboratory developed tests/experiments, specimen engraftment potency, and use of leucoreduced blood products for transfusion.
BD FACSLyric Flow Cytometer, 2-Laser 6 Color Instrument, REF: 662876, when using BD Trucount Tubes
Becton, Dickinson and Company, BD Biosciences
Flow cytometers, when used with specific tubes may experience excessive electronic abort counts (electronic abort count of beads exceeds 1.0% of the total processed events), which may affect the ratio of cell population events to bead events, potentially resulting in falsely high absolute counts. This may affect the following: initiation of prophylactic therapy for Opportunistic Infections (HIV patients), decisions regarding care/management (PID/immune suppressed patients), laboratory developed tests/experiments, specimen engraftment potency, and use of leucoreduced blood products for transfusion.
SPIFE ImmunoFix-6 Kit, Cat. No. 3401, 3401T, For In-Vitro Diagnostic Use. The firm name on the label is Helena Laboratories, Beaumont, TX.
Helena Laboratories
The plates do not have wells punched for the controls in the patient 1 area.
Achilles Insight Bone Sonometer
GE Healthcare
Certain Achilles Express and Achilles Insight systems were shipped with European CEE-7/7 Type Power Cords to countries where three protruding pin plugs are required for protective earth connection.
Aperio CS2 CE IVD scanner instrument, scanner part #23CS100CE.
Leica Biosystems Imaging
The incorrect sensor was used in the assembly of the camera in the scanner instrument which can potentially generate whole slide images that impact analysis algorithm results, as well as potentially impact the visual appearance of whole slide images.
BD FACSLyric Flow Cytometer,3-Laser 10 Color Instrument, REF: 659180 and 65918001, when using BD Trucount Tubes
Becton, Dickinson and Company, BD Biosciences
Flow cytometers, when used with specific tubes may experience excessive electronic abort counts (electronic abort count of beads exceeds 1.0% of the total processed events), which may affect the ratio of cell population events to bead events, potentially resulting in falsely high absolute counts. This may affect the following: initiation of prophylactic therapy for Opportunistic Infections (HIV patients), decisions regarding care/management (PID/immune suppressed patients), laboratory developed tests/experiments, specimen engraftment potency, and use of leucoreduced blood products for transfusion.
The action is being initiated because the cooling indicator board was improperly installed. This was first noticed by a service employee during assembly. If for some reason the cooling of the reagent rotor fails, the reagent cooling indicator light will not illuminate and the acoustic beeper will not go off. As a consequence the operator will not be warned and the reagent storage conditions may fall out of specification.