Online TDM Vancomycin Gen.3 ( 200 tests) Catalog # 06779344190 Analyzer/Module: cobas c 311 and cobas c 501/502
Roche Diagnostics Operations
Inaccurately Low Sample Results
π₯ Medical Devices β’ 6,627 recalls
Roche Diagnostics Operations
Inaccurately Low Sample Results
Becton, Dickinson and Company, BD Biosciences
Flow cytometers, when used with specific tubes may experience excessive electronic abort counts (electronic abort count of beads exceeds 1.0% of the total processed events), which may affect the ratio of cell population events to bead events, potentially resulting in falsely high absolute counts. This may affect the following: initiation of prophylactic therapy for Opportunistic Infections (HIV patients), decisions regarding care/management (PID/immune suppressed patients), laboratory developed tests/experiments, specimen engraftment potency, and use of leucoreduced blood products for transfusion.
Becton, Dickinson and Company, BD Biosciences
Flow cytometers, when used with specific tubes may experience excessive electronic abort counts (electronic abort count of beads exceeds 1.0% of the total processed events), which may affect the ratio of cell population events to bead events, potentially resulting in falsely high absolute counts. This may affect the following: initiation of prophylactic therapy for Opportunistic Infections (HIV patients), decisions regarding care/management (PID/immune suppressed patients), laboratory developed tests/experiments, specimen engraftment potency, and use of leucoreduced blood products for transfusion.
Meters may contain a damaged C29 capacitor resulting in issues with power allegation (short battery life or meter not turning on).
Stryker Instruments Div. of Stryker
Stryker C2 NerveMonitor's design and instructions may not optimally address usability issues related to the functionality of the device, which may result in use errors potentially causing or contributing to nerve injury.
WOM World of Medicine AG
The deficit displayed by the pump can differ from the real deficit, leading potentially to patient risk.
Meters may contain a damaged C29 capacitor resulting in issues with power allegation (short battery life or meter not turning on).
The touchpad may not work properly after cleaning.
Stryker Instruments Div. of Stryker
Stryker C2 NerveMonitor's design and instructions may not optimally address usability issues related to the functionality of the device, which may result in use errors potentially causing or contributing to nerve injury.
Software malfunction; The action being taken due to the device becoming inoperable due to software lockup.
Helena Laboratories
Some of the cards show occasional shorter clot times than the expected variation resulting in a potential false negative result on the screening assay.
Randox Laboratories
Analyte range for Leukocytes for use with the Siemens Multistix method is incorrect. The range was corrected to NEGATIVE to 3+.
Atellica IM anti-CCP IgG (aCCP) Test Definition (TDef) Mitigation Failure-reagent carryover mitigation for the Testosterone II (TSTII) assay is unable to be completed as defined in the aCCP Test Definition (TDef). The issue occurs on an Atellica IM Analyzer and is isolated to TDef Version 1.0. FOR OUS: Atellica IM anti-CCP IgG (aCCP) and Pregnancy-Associated Plasma Protein-A (PAPP-A) Test Definition (TDef) Mitigation Failure
Instrument stopped working due to a software lockup, and no patient results are produced.
The electrode lead connector has a protruding conductor beyond insulation of approximately 0.1 mm.
Remote Antenna Cable Used with Philips MX4O Monitors Does Not Have the Required Rating for Some Installations
Non-conforming devices are identified, which may not have been included in prior field actions and may not have had a number of corrections applicable to these devices.
Philips North America
The thickness of the connector to the trunk cable on affected limb lead sets may be greater than specified. which may prevent or impede the performance of I 2-Lead ECG measurements.
Randox Laboratories realigned RX Analyser Series calibration targets by 10% for AST and 9% for ALT in line with the Mean of All Instrument method target for the Randox Calibration Serum Level 3 (for calibrator lot 961UE onwards).
Stryker Communications
The mounts on the monitor may have an insufficient weld.