It has been determined that certain Convenience kits that were packaged in breather pouch lot 323080002 may result in the potential of using non-sterile product if the vendor seal of the pouch is opened/unsealed.
Other Medical Devices
🏥 Medical Devices • 15,809 recalls
Medline brand medical procedure convenience kit, labeled as: a) PAIN PROCEDURE PACK, REF DYNJ80370A; b) DR HEFZY BLOCK TRAY, REF DYNJRA1854A
MEDLINE INDUSTRIES, LP - Northfield
It has been determined that certain Convenience kits that were packaged in breather pouch lot 323080002 may result in the potential of using non-sterile product if the vendor seal of the pouch is opened/unsealed.
smiths medical portex, Thermovent 1200, 15mm/22mm: a) REF 100/582/000; b) REF 100/582/000JP; c) REF 100/582/000JS
Smiths Medical ASD
Uncertainty in the seal integrity of the sterile packaging.
Medline DIALYSIS PACK, REF DYNJ36932A; medical procedure convenience kit
MEDLINE INDUSTRIES, LP - Northfield
It has been determined that certain Convenience kits that were packaged in breather pouch lot 323080002 may result in the potential of using non-sterile product if the vendor seal of the pouch is opened/unsealed.
medex TranStar KIDS KIT 7INCH CLOSED BLOOD SAMPLE KIT with 30cc/hr Flush Device 10/EA, Product Code REF MX9547
Smiths Medical ASD
Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.
Baxter Operating Table TruSystem 7000 V, Product code REF 1841050
Baxter Healthcare
The firm received customer reports stating that the batteries and their connectors experienced electrical short-circuits and/or emitted smoke.
ENT RES 100 ML Non-Sterile W CLAMP, TUBING FLAG, & FEMALE LUER, Product Code 68-2230-28
Smiths Medical ASD
Certain CADD Medication Cassette Reservoirs may exhibit a weakened weld joint between the medication bag and tubing due to a production equipment malfunction. This could result in a medication leakage.
Medline TUBAL LIGATION PACK, REF DYNJ66230B; medical procedure convenience kit
MEDLINE INDUSTRIES, LP - Northfield
It has been determined that certain Convenience kits that were packaged in breather pouch lot 323080002 may result in the potential of using non-sterile product if the vendor seal of the pouch is opened/unsealed.
medex TranStar KIDS KIT 34IN. CLOSED BLOOD SAMPLE KIT with 30cc/hr Flush Device 10/EA, Product Code REF MX9545
Smiths Medical ASD
Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.
Baxter Operating Table TruSystem 7000, Product code REF 1841046
Baxter Healthcare
The firm received customer reports stating that the batteries and their connectors experienced electrical short-circuits and/or emitted smoke.
Centurion PEDIATRIC/ADULT TRACHEOSTOMY T, REF TC7895; medical procedure convenience kit
MEDLINE INDUSTRIES, LP - Northfield
It has been determined that certain Convenience kits that were packaged in breather pouch lot 323080002 may result in the potential of using non-sterile product if the vendor seal of the pouch is opened/unsealed.
medex TranStar KIDS KIT 10IN. CLOSED BLOOD SAMPLE KIT with 30cc/hr Flush Device 10/EA, Product Code REF MX9548
Smiths Medical ASD
Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.
medex TranStar DPT w/Two Stopcocks 10/EA, Product Code REF MX950X2SC
Smiths Medical ASD
Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.
Baxter Operating Table TruSystem 7000 (dV), Product code REF 1841049
Baxter Healthcare
The firm received customer reports stating that the batteries and their connectors experienced electrical short-circuits and/or emitted smoke.
AD-TECH Spencer Probe Depth Electrode, REF SD04R-AP58X-000 4-Contact 1.12mm (A style) Depth Electrode
Ad-Tech Medical Instrument
Ad-Tech is recalling these A-Style SD Depth Electrodes because a label discrepancy was identified on A style contact SD electrodes where the MR Conditional symbol was incorrectly placed instead of the MR Unsafe symbol. The A style contacts were not cleared for MR Conditional use. Our records indicate you have received one or more of the affected lots listed above.
Product has been identified as having holes in the packaging.
medex TranStar SINGLE 84IN SET WITH ANTI-SHUNT STOPCOCK 10/EA, Product Code REF MXG300243R1
Smiths Medical ASD
Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.
Baxter Operating table TruSystem 7000 U (MB), Product codes: a) REF 1604786; b) REF 1604788; c) REF 1723633
Baxter Healthcare
The firm received customer reports stating that the batteries and their connectors experienced electrical short-circuits and/or emitted smoke.
Certain CADD Medication Cassette Reservoirs may exhibit a weakened weld joint between the medication bag and tubing due to a production equipment malfunction. This could result in a medication leakage.
Medline brand medical procedure convenience kits, labeled as: a) ABDOMINAL HYSTERECTOMY TEGH-LF, REF DYNJ46730; b) HYSTEROSCOPY PACK, REF DYNJ49302A; c) PERI GYN PACK, REF DYNJ57526A; d) HYSTEROSCOPY TRAY ACS, REF DYNJ68750
MEDLINE INDUSTRIES, LP - Northfield
It has been determined that certain Convenience kits that were packaged in breather pouch lot 323080002 may result in the potential of using non-sterile product if the vendor seal of the pouch is opened/unsealed.