It has been determined that certain Convenience kits that were packaged in breather pouch lot 323080002 may result in the potential of using non-sterile product if the vendor seal of the pouch is opened/unsealed.
Other Medical Devices
🏥 Medical Devices • 15,809 recalls
CADD Medication Cassette Reservoir, 100mL, FS, Product Code 21-7302-24
Smiths Medical ASD
Certain CADD Medication Cassette Reservoirs may exhibit a weakened weld joint between the medication bag and tubing due to a production equipment malfunction. This could result in a medication leakage.
Baxter Operating Table TruSystem 7000 (MBW), Product code REF 1841048
Baxter Healthcare
The firm received customer reports stating that the batteries and their connectors experienced electrical short-circuits and/or emitted smoke.
CADD Medication Cassette Reservoir, 50mL, FS, Product Code 21-7301-24
Smiths Medical ASD
Certain CADD Medication Cassette Reservoirs may exhibit a weakened weld joint between the medication bag and tubing due to a production equipment malfunction. This could result in a medication leakage.
medex TranStar Single ICP Set 1/EA, Product Code REF MX20897
Smiths Medical ASD
Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.
medex TranStar Single OR Set 1/EA, Product Code REF MX20846R2
Smiths Medical ASD
Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.
AD-TECH Spencer Probe Depth Electrode, REF SD06R-AP58X-000 6-Contact 1.12mm (A style) Depth Electrode
Ad-Tech Medical Instrument
Ad-Tech is recalling these A-Style SD Depth Electrodes because a label discrepancy was identified on A style contact SD electrodes where the MR Conditional symbol was incorrectly placed instead of the MR Unsafe symbol. The A style contacts were not cleared for MR Conditional use. Our records indicate you have received one or more of the affected lots listed above.
Medline brand, medical procedure convenience kits, labeled as: a) SET UP PACK, REF DYNJ36320A; b) ANGIOGRAM WRO-LF, REF DYNJ41876B
MEDLINE INDUSTRIES, LP - Northfield
It has been determined that certain Convenience kits that were packaged in breather pouch lot 323080002 may result in the potential of using non-sterile product if the vendor seal of the pouch is opened/unsealed.
Amingo. OUS only.
Datex-Ohmeda
Potential issue that may arise if the limb of a patient breathing circuit is mistakenly connected to the Auxiliary Common Gas Outlet (ACGO) port.
It has been determined that certain Convenience kits that were packaged in breather pouch lot 323080002 may result in the potential of using non-sterile product if the vendor seal of the pouch is opened/unsealed.
CADD Medication Cassette Reservoir, 50mL, Product Code 21-7001-24
Smiths Medical ASD
Certain CADD Medication Cassette Reservoirs may exhibit a weakened weld joint between the medication bag and tubing due to a production equipment malfunction. This could result in a medication leakage.
Medline brand medical procedure convenience kits, labeled as: a) NEURO PACK, REF DYNJ37909A; b) SHUNT PACK, REF DYNJ40040C
MEDLINE INDUSTRIES, LP - Northfield
It has been determined that certain Convenience kits that were packaged in breather pouch lot 323080002 may result in the potential of using non-sterile product if the vendor seal of the pouch is opened/unsealed.
CADD Medication Cassette Reservoir, 50mL, FS, Product Code 21-7301-24JP
Smiths Medical ASD
Certain CADD Medication Cassette Reservoirs may exhibit a weakened weld joint between the medication bag and tubing due to a production equipment malfunction. This could result in a medication leakage.
medex TranStar Kit w/ Stopcock1/EA, Product Code REF MX20740
Smiths Medical ASD
Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.
Medline PEDIATRIC DRAPE PK, REF DYNJ81297; medical procedure convenience kit
MEDLINE INDUSTRIES, LP - Northfield
It has been determined that certain Convenience kits that were packaged in breather pouch lot 323080002 may result in the potential of using non-sterile product if the vendor seal of the pouch is opened/unsealed.
It has been determined that certain Convenience kits that were packaged in breather pouch lot 323080002 may result in the potential of using non-sterile product if the vendor seal of the pouch is opened/unsealed.
Due to labeling issue. The box label is different that the gallon label in that gallon label shows a lower Calcium concentration than the box label.
medex TranStar KIDS KIT 19IN. CLOSED BLOOD SAMPLE KIT with 30cc/hr Flush Device 10/EA, Product Code REF MX9543
Smiths Medical ASD
Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.
medex TranStar Disposable Transducer 1/EA, Product Code REF MX950
Smiths Medical ASD
Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.
Medline CV ANESTHESIA - ROOM SET UP, REF DYNJ905503F; medical procedure convenience kit
MEDLINE INDUSTRIES, LP - Northfield
It has been determined that certain Convenience kits that were packaged in breather pouch lot 323080002 may result in the potential of using non-sterile product if the vendor seal of the pouch is opened/unsealed.