Haemonetics has received isolated reports that in some cases the 70ml centrifuge bowl may leak during cell salvage processing.
Other Medical Devices
🏥 Medical Devices • 15,809 recalls
COOK MEDICAL Captura SERRATED BIOPSY FORCEPS WITH SPIKE,2.4 mm, 230 cm, with Spike, DBF-2.4SN-230SP-20-S
Wilson-Cook Medical
The product is being recalled because Captura Serrated Forceps with Spike (DBF-2.4SN-230SP-20-S) were mislabeled as Captura Serrated Large Forcep-No Spike (DBF-2.4SL-230-20-S) and Captura Serrated Large Forcep-No Spike (DBF-2.4SL-230-20-S) were mislabeled as Captura Serrated Forceps with Spike (DBF-2.4SN-230SP-20-S).
SUCTION / ANTICOAGULATION ASSEMBLY (a) BTC93 (b) BTC98 Suction and Anticoagulation assemblies are indicated for use with an autotransfusion system between the suction tip and reservoir.
Medtronic Perfusion Systems
Medtronic discovered that certain Y-connectors included in the suction lines have exhibited tearing and cracking. These defects could cause the Y-connectors to leak when in use, which could result in potential blood loss and a breach in the circuit.
COOK MEDICAL Captura SERRATED LARGE FORCEP-NO SPIKE, 2.4 mm, 230 cm, DBF-2.4SL- 230-20-S
Wilson-Cook Medical
The product is being recalled because Captura Serrated Forceps with Spike (DBF-2.4SN-230SP-20-S) were mislabeled as Captura Serrated Large Forcep-No Spike (DBF-2.4SL-230-20-S) and Captura Serrated Large Forcep-No Spike (DBF-2.4SL-230-20-S) were mislabeled as Captura Serrated Forceps with Spike (DBF-2.4SN-230SP-20-S).
Vial2BAG DC 13mm, REF 6070111, Single use only, Rx Only
WEST PHARMA. SERVICES IL
Adverse events have been reported with the use of oxytocin in conjunction with the product in pregnant women. There is a possibility that the device may cause inadequate admixing of drug with the solution in the IV bag. This could lead to concentrated drug pooling in the device and bolus injections.
Vial2BAG DC 20mm: 1) REF 6070112 & 2) REF 6070112C, Single use only, Rx Only
WEST PHARMA. SERVICES IL
Adverse events have been reported with the use of oxytocin in conjunction with the 13mm Vial2Bag DC in pregnant women. Vial2Bag DC 20mm is recalled due to similarities with 13mm Vial2Bag DC.
Haemonetics has received isolated reports that in some cases the 70ml centrifuge bowl may leak during cell salvage processing.
Vial2BAG Transfer Device, REF 6070104, Single use only, Rx Only
WEST PHARMA. SERVICES IL
Adverse events have been reported with the use of oxytocin in conjunction with the 13mm Vial2Bag DC in pregnant women. Vial2Bag is recalled due to similarities with 13mm Vial2Bag DC.
RESERVOIR "Y" ADAPTER, (a) BT920 (b) BT926 Suction and Anticoagulation assemblies are indicated for use with an autotransfusion system between the suction tip and reservoir.
Medtronic Perfusion Systems
Medtronic discovered that certain Y-connectors included in the suction lines have exhibited tearing and cracking. These defects could cause the Y-connectors to leak when in use, which could result in potential blood loss and a breach in the circuit.
Heating Profile PRISMAFLO II S blood return warmer for Prismaflex Product Code: PF2-WP31 Product Usage: The PRISMAFLO IIS warmer is used for warming return blood flow.
Stihler Electronic Gmbh
PRISMAFLO II and PRISMAFLO II S Heating Profile only and not the Control Unit, damage may result in a hotspot on the Heating Profile, may cause low level skin burn
There is an anomaly with the Eclipse(TM) Treatment Planning System [TPS] Proton Convolution Superposition [PCS] dose calculation algorithm. The PCS algo- rithm calculates the water equivalent range incorrectly for non-square 3D CT images (either different number of pixels in X and Y, or non-square pixels). Before the dose is calculated, Eclipse resamples the CT images to create a calculation image with a maximum resolution of 256 x 256. The PCS algorithm assumes uniform resolution in X and Y directions, or that X = Y for all images, and erroneously sets the Y = X for dose calcu- lation. For images that are not square, where either X<>Y, or the length of X does not equal the length of Y, the computed water equivalent range R is erroneous compared to the correct range R: 1) X=Y; R =R (R is correct); 2) X<Y: R <R (R is too small) and 3) X>Y R >R (R is too big).
Heating Profile PRISMAFLO II S blood return warmer for Prismaflex Product Code: PF2-WP33 Product Usage: The PRISMAFLO IIS warmer is used for warming return blood flow.
Stihler Electronic Gmbh
PRISMAFLO II and PRISMAFLO II S Heating Profile only and not the Control Unit, damage may result in a hotspot on the Heating Profile, may cause low level skin burn
Perfusion Pack, Rx only, Sterile EO
LivaNova USA
4-way stopcocks and stopcock manifold assembled in certain perfusion tubing systems, indicated for use in surgical procedures requiring extracorporeal support for periods of up to six hours, may leak or be occluded.
Heating Profile PRISMAFLO II blood return warmer for Prismaflex Product Code: PF2-WP2618 Product Usage: The PRISMAFLO IIS warmer is used for warming return blood flow.
Stihler Electronic Gmbh
PRISMAFLO II and PRISMAFLO II S Heating Profile only and not the Control Unit, damage may result in a hotspot on the Heating Profile, may cause low level skin burn
U by Kotex¿ Sleek¿, Regular Tampons, 3 Count
Kimberly-Clark
Potential for tampon to come apart upon removal.
U by Kotex¿ Sleek¿, Regular Tampons, 18 Count
Kimberly-Clark
Potential for tampon to come apart upon removal.
Porcine Anorganic Bone Mineral 4.0cc
Collagen Matrix
There is a possibility that the product labeled as 4.0cc volume may only contain 2.0cc volume and is therefore mislabeled.
The pin that holds the spring mechanism is not properly mounted. The pin could potentially disassemble.
Potential for tampon to come apart upon removal.
If the head mount knob is not properly tightened and the operator directly uses the injector head to move the unit, it may become disengaged, causing the injector head to become loose or fall.