Other Medical Devices

🏥 Medical Devices 15,809 recalls

Haemonetics has received isolated reports that in some cases the 70ml centrifuge bowl may leak during cell salvage processing.

Aug 2, 2017 Other Medical Devices Nationwide View Details →

The product is being recalled because Captura Serrated Forceps with Spike (DBF-2.4SN-230SP-20-S) were mislabeled as Captura Serrated Large Forcep-No Spike (DBF-2.4SL-230-20-S) and Captura Serrated Large Forcep-No Spike (DBF-2.4SL-230-20-S) were mislabeled as Captura Serrated Forceps with Spike (DBF-2.4SN-230SP-20-S).

Dec 18, 2018 Other Medical Devices View Details →

Medtronic discovered that certain Y-connectors included in the suction lines have exhibited tearing and cracking. These defects could cause the Y-connectors to leak when in use, which could result in potential blood loss and a breach in the circuit.

Nov 27, 2018 Other Medical Devices Nationwide View Details →

The product is being recalled because Captura Serrated Forceps with Spike (DBF-2.4SN-230SP-20-S) were mislabeled as Captura Serrated Large Forcep-No Spike (DBF-2.4SL-230-20-S) and Captura Serrated Large Forcep-No Spike (DBF-2.4SL-230-20-S) were mislabeled as Captura Serrated Forceps with Spike (DBF-2.4SN-230SP-20-S).

Dec 18, 2018 Other Medical Devices View Details →
Class I - Dangerous

Adverse events have been reported with the use of oxytocin in conjunction with the product in pregnant women. There is a possibility that the device may cause inadequate admixing of drug with the solution in the IV bag. This could lead to concentrated drug pooling in the device and bolus injections.

Dec 26, 2018 Other Medical Devices View Details →

Haemonetics has received isolated reports that in some cases the 70ml centrifuge bowl may leak during cell salvage processing.

Aug 2, 2017 Other Medical Devices Nationwide View Details →

Medtronic discovered that certain Y-connectors included in the suction lines have exhibited tearing and cracking. These defects could cause the Y-connectors to leak when in use, which could result in potential blood loss and a breach in the circuit.

Nov 27, 2018 Other Medical Devices Nationwide View Details →

There is an anomaly with the Eclipse(TM) Treatment Planning System [TPS] Proton Convolution Superposition [PCS] dose calculation algorithm. The PCS algo- rithm calculates the water equivalent range incorrectly for non-square 3D CT images (either different number of pixels in X and Y, or non-square pixels). Before the dose is calculated, Eclipse resamples the CT images to create a calculation image with a maximum resolution of 256 x 256. The PCS algorithm assumes uniform resolution in X and Y directions, or that X = Y for all images, and erroneously sets the Y = X for dose calcu- lation. For images that are not square, where either X<>Y, or the length of X does not equal the length of Y, the computed water equivalent range R is erroneous compared to the correct range R: 1) X=Y; R =R (R is correct); 2) X<Y: R <R (R is too small) and 3) X>Y R >R (R is too big).

Dec 26, 2018 Other Medical Devices Nationwide View Details →
Class I - Dangerous

4-way stopcocks and stopcock manifold assembled in certain perfusion tubing systems, indicated for use in surgical procedures requiring extracorporeal support for periods of up to six hours, may leak or be occluded.

Nov 9, 2018 Other Medical Devices Nationwide View Details →

If the head mount knob is not properly tightened and the operator directly uses the injector head to move the unit, it may become disengaged, causing the injector head to become loose or fall.

Dec 14, 2018 Other Medical Devices Nationwide View Details →