The recall was initiated because Beckman Coulter has confirmed that some AU Clinical Chemistry Analyzer ISE units can experience premature ISE REF (Reference) valve failure. Premature valve failure can be detected by the presence of air bubbles in the ISE REF solution tubing on the ISE unit. If air bubbles are present, accuracy of the ISE test results may be affected.

Sep 9, 2013 Other Medical Devices Nationwide View Details β†’

Stryker has identified a potential for the auger and/or cannula to fracture when the Dekompressor is used contrary to the warnings in the Instructions For Use (IFU). Specifically, manually bending the cannula can weaken the device and may result in a fracture of the auger and/or cannula. This could result in the need for additional medical intervention to remove the broken piece, pain or loss of m

Nov 4, 2013 Other Medical Devices Nationwide View Details β†’

The snap-on fixtures of the backrest cushion on the Mobile Shower and Toilet Commode Aquatec Ocean VIP and Ocean Dual VIP do not withstand a required pull out force of 500 N, when the product is used in a reclined position and or tilted position. This may result in the snap-on fixture coming unlatched.

Aug 31, 2013 Other Medical Devices View Details β†’

Siemens has discovered through product monitoring a potential malfunction when using system operating with software versions VB10D, VB10F, VB10C and VB10. During a RAD examination the possibility exists that radiation may continue longer than expected and may result in unnecessary dose applied during an acquisition. Additionally, it is not possible for the user to stop the radiation manually; ho

May 6, 2013 Other Medical Devices Nationwide View Details β†’

Edwards Lifesciences is recalling two lots of Duraflo Cannulas due to reported complaints involving cracks in the t-connector of the devices which were confirmed during product evaluation.

Aug 26, 2013 Other Medical Devices Nationwide View Details β†’

Xoran Minicat00158

Xoran Technologies

Class I - Dangerous

The MiniCAT CT Scanners subject to this notification failed to meet Xoran's specifications for dose rate of 125 ΒΏ 3 kVp. These 18 complaints received include dose rates both on the high and low end of Xoran's specifications. See ATTACHMENT 1 for details. (e) A n evaluation of the hazards reasonably related to defect or the failure to comply with the Federal standard; An evaluation of the health ha

Jul 27, 2010 Other Medical Devices Nationwide View Details β†’

During an audit, Carestream discovered a DRX-1 System that captured an image which displayed an artifact. This issue was discovered in manufacturing. It occurs when the DRX product console software is updated to v5.6b software and the detector is not calibrated before use.

Jul 26, 2013 Other Medical Devices Nationwide View Details β†’

The sterility of this product to a Sterility Assurance Level (SAL) of 10(-6) cannot be assured.

Aug 28, 2013 Other Medical Devices Nationwide View Details β†’