Xoran Technologies

Class I - Dangerous

The MiniCAT CT Scanners subject to this notification failed to meet Xoran's specifications for dose rate of 125 ΒΏ 3 kVp. These 18 complaints received include dose rates both on the high and low end of Xorans specifications.

Jul 27, 2010 Other Medical Devices Nationwide View Details β†’

The low density polyethylene (LDPE) bag used to package implants adheres to the highly polished surface of the devices.

Aug 30, 2013 Other Medical Devices Nationwide View Details β†’

Edwards Lifesciences is recalling certain lots of EMBOL-X Glide Protection System due to deformation of the cannula tip.

Sep 16, 2013 Other Medical Devices Nationwide View Details β†’

NexGenΒΏ Femoral Component (CR-FLEX GSF PRECOAT SZ C-L; CR-FLEX GSF PRECOAT SZ C-R ; CR-FLEX GSF PCT SZ C-L MINUS; CR-FLEX GSF PCT SZ C-R MINUS ;CR-FLEX GSF PRECOAT SZ D-L; CR-FLEX GSF PRECOAT SZ D-R ;CR-FLEX GSF PCT SZ D-L MINUS ; CR-FLEX GSF PCT SZ D-R MINUS ;CR-FLEX GSF PRECOAT SZ E-L CR-FLEX GSF PRECOAT SZ E-R ;CR-FLEX GSF PCT SZ E-L MINUS ;CR-FLEX GSF PCT SZ E-R MINUS ;CR-FLEX GSF PRECOAT SZ F-L ;CR-FLEX GSF PRECOAT SZ F-R ;CR-FLEX GSF PCT SZ F-L MINUS ;CR-FLEX GSF PCT SZ F-R MINUS ;CR-FLEX GSF PRECOAT SZ G-L ;CR-FLEX GSK PRECOAT SZ G-R ;CR-FLEX GSF PCT SZ G-L MINUS ;CR-FLEX GSF PCT SZ G-R MINUS ; LPS-FLEX GSF OPT SZ C-L ; LPS-FLEX GSF OPT SZ C-R ;LPS-FLEX GSF OPT SZ D-L ;LPS-FLEX GSF OPT SZ D-R ;LPS-FLEX GSF OPT SZ E-L ;LPS-FLEX GSF OPT SZ E-R ;LPS-FLEX GSK OPT SZ F-L ;LPS-FLEX GSF OPT SZ F-R ;LPS-FLEX GSF OPT SZ G-L ; LPS-FLEX GSF OPT SZ G-R ;CR-FLEX PCT FEM B-L; CR-FLEX PCT FEM B-R; CR-FLEX PCT FEM C-L; CR-FLEX PCT FEM C-R; CR-FLEX PCT FEM C-L MINUS; CR-FLEX PCT FEM C-R MINUS; CR-FLEX PCT FEM D-L; CR-FLEX PCT FEM D-R; CR-FLEX PCT FEM D-L MINUS; CR-FLEX PCT FEM D-R MINUS; CR-FLEX PCT FEM E-L; CR-FLEX PCT FEM E-R; CR-FLEX PCT FEM E-L MINUS; CR-FLEX PCT FEM E-R MINUS; CR-FLEX PCT FEM F-L; CR-FLEX PCT FEM F-R; CR-FLEX PCT FEM F-L MINUS; CR-FLEX PCT FEM F-R MINUS; CR-FLEX PCT FEM G-L; CR-FLEX PCT FEM G-R; CR-FLEX PCT FEM G-L MINUS; CR-FLEX PCT FEM G-R MINUS; CR-FLEX POR FEM B-L; CR-FLEX POR FEM B-R; CR-FLEX POR FEM C-L; CR-FLEX POR FEM C-R; CR-FLEX POR FEM C-L MINUS; CR-FLEX POR FEM C-R MINUS; CR-FLEX POR FEM D-L; CR-FLEX POR FEM D-R; CR-FLEX POR FEM D-L MINUS; CR-FLEX POR FEM D-R MINUS; CR-FLEX POR FEM E-L; CR-FLEX POR FEM E-R; CR-FLEX POR FEM E-L MINUS; CR-FLEX POR FEM E-R; MINUS; CR-FLEX POR FEM F-L; CR-FLEX POR FEM F-R ;CR-FLEX POR FEM F-L MINUS; CR-FLEX POR FEM F-R MINUS; CR-FLEX POR FEM G-L; CR-FLEX POR FEM G-R; CR-FLEX POR FEM G-L MINUS CR-FLEX POR FEM G-R MINUS; CR-FLEX OPT FEM B-L; CR-FLEX OPT FEM B-R; CR-FLEX OPT FEM C-L; CR-FLEX OPT FEM C-R; CR-FLEX OPT FEM C-L MINUS; CR-FLEX OPT FEM C-R MINUS; CR-FLEX OPT FEM D-L; CR-FLEX OPT FEM D-R; CR-FLEX OPT FEM D-L MINUS; CR-FLEX OPT FEM D-R MINUS; CR-FLEX OPT FEM E-L; CR-FLEX OPT FEM E-R; CR-FLEX OPT FEM E-L MINUS; CR-FLEX OPT FEM E-R MINUS; CR-FLEX OPT FEM F-L; CR-FLEX OPT FEM F-R; CR-FLEX OPT FEM F-L MINUS; CR-FLEX OPT FEM F-R MINUS; CR-FLEX OPT FEM G-L; CR-FLEX OPT FEM G-R; CR-FLEX OPT FEM G-L MINUS; CR-FLEX OPT FEM G-R MINUS; LPS-FLEX CO-NID FEM A-L ; LPS-FLEX CO-NID FEM A-R ; LPS-FLEX RECOAT FEMORAL A-L; LPS-FLEX PRECOAT FEMORAL A-R; LPS-FLEX CO-NID FEM B-L ; LPS-FLEX CO-NID FEM B-R ; LPS-FLEX PRECOAT FEMORAL B-L; LPS-FLEX PRECOAT FEMORL B-R; LPS-FLEX CO-NID FEM C-L ; LPS-FLEX CO-NID FEM C-R ; LPS-FLEX PRECOAT FEMORAL C-L; LPS-FLEX PRECOAT FEMORAL C-R; LPS-FLEX CO-NID FEM D-L ; LPS-FLEX CO-NID FEM D-R ; LPS-FLEX PRECOAT FEMORAL D-L; LPS-FLEX PRECOAT FEMORAL D-R; LPS-FLEX CO-NID FEM E-L ; LPS-FLEX CO-NID FEM E-R ; LPS-FLEX PRECOAT FEMORAL E-L; LPS-FLEX PRECOAT FEMORAL E-R; LPS-FLEX CO-NID FEM F-L ; LPS-FLEX CO-NID FEM F-R ; LPS-FLEX PRECOAT FEMORAL F-L; LPS-FLEX PRECOAT FEMORAL F-R; LPS-FLEX CO-NID FEM G-L ; LPS-FLEX CO-NID FEM G-R ; LPS-FLEX PRECOAT FEMORAL G-L; LPS-FLEX PRECOAT FEMORAL G-R; LPS-FLEX PRECOAT FEMORAL H-L; LPS-FLEX PRECOAT FEMORAL H-R ; LPS-FLEX POROUS FEMORAL B-L ; LPS-FLEX POROUS FEMORAL B-R ; LPS-FLEX POROUS FEMORAL C-L ; LPS-FLEX POROUS FEMORAL C-R LPS-FLEX POROUS FEMORAL D-L ; LPS-FLEX POROUS FEMORAL D-R ; LPS-FLEX POROUS FEMORAL E-L ; LPS-FLEX POROUS FEMORAL E-R ; LPS-FLEX POROUS FEMORAL F-L ; LPS-FLEX POROUS FEMORAL F-R ;LPS-FLEX POROUS FEMORAL G-L ; LPS-FLEX POROUS FEMORAL G-R ; LPS-FLEX POROUS FEMORAL H-L ; LPS-FLEX POROUS FEMORAL H-R ; LPS-FLEX OPTION FEMORAL A-L; LPS-FLEX OPTION FEMORAL A-R; LPS-FLEX OPTION FEMORAL B-L; LPS-FLEX OPTION FEMORAL B-R; LPS-FLEX OPTION FEMORAL B-L ; LPS-FLEX OPTION FEMORAL B-R ; LPS-FLEX OPTION FEMORAL C-L; LPS-FLEX OPTION FEMORAL C-R; LPS-FLEX OPTION FEMORAL C-L; LPS-FLEX OPTION FEMORAL C-R; LPS-FLEX OPTI

Zimmer

Class I - Dangerous

The low density polyethylene (LDPE) bag used to package implants adheres to the highly polished surface of the devices.

Aug 30, 2013 Other Medical Devices Nationwide View Details β†’

It has come to the attention of MAQUET that in certain rare instances, the CARDIOHELP-i device may display an error message with audible alarm, indicating "battery defective." The alarm will occur-upon start up or when the main power supply is unplugged from the main power source. The human machine interface touch screen of the CARDIOHELP-i device may gradually become black due to the human machine interface (HMI) back light shutting down.

Sep 12, 2013 Other Medical Devices Nationwide View Details β†’

Xoran Minicat00202

Xoran Technologies

Class I - Dangerous

The MiniCAT CT Scanners subject to this notification failed to meet Xoran's specifications for dose rate of 125 ΒΏ 3 kVp. These 18 complaints received include dose rates both on the high and low end of Xoran's specifications. See ATTACHMENT 1 for details. (e) A n evaluation of the hazards reasonably related to defect or the failure to comply with the Federal standard; An evaluation of the health ha

Jul 27, 2010 Other Medical Devices Nationwide View Details β†’

An in-vitro diagnostic reagent may be labeled with an incorrect barcode, which when scanned on associated diagnostic instrument may cause a false negative result.

Sep 9, 2013 Other Medical Devices Nationwide View Details β†’

The recall was initiated after Varian Medical became aware the Lexan templates of the Breast Bridge Template System could become bent and turn opaque after steam sterilization.

Aug 9, 2013 Other Medical Devices Nationwide View Details β†’

OPTI LION E-Plus Cassettes were labeled with an incorrect barcode. The incorrect barcode could perfect the performance of one or more parameters as the product ages. Also a Customer complaint from a distributor reported that a customer was getting QC failures.

Aug 9, 2013 Other Medical Devices Nationwide View Details β†’

Some devices may be contaminated on the back side of the device with glue in the blood application area.

Sep 5, 2013 Other Medical Devices View Details β†’

Siemens determined that if an autosampler rack jam error occurs during operation of an ADVIA 2120/2120i connected to an ADVIA Autoslide, it is possible that the next slide processed by the Autoslide could be labeled with the wrong sample identification information.

Aug 12, 2013 Other Medical Devices Nationwide View Details β†’

GE Healthcare has recently become aware of a potential safety issue related with the DMPR function and Screen Saved Images on your scanner which involves Optima CT660, Brivo CT385, Discovery PET/CT 610, Discovery PET/CT 710. The DMPR and Screen Save functionality may possibly generate the same Series UID. It has been seen that some PACS may combine series under a patient using only the Series UID

Mar 26, 2013 Other Medical Devices Nationwide View Details β†’