Dilators and Holding Clip are being used for a secondary function of neuromonitoring, for which it is not 510(k) cleared.
Other Medical Devices
π₯ Medical Devices β’ 15,809 recalls
Xoran MiniCAT00202
Xoran Technologies
The MiniCAT CT Scanners subject to this notification failed to meet Xoran's specifications for dose rate of 125 ΒΏ 3 kVp. These 18 complaints received include dose rates both on the high and low end of Xorans specifications.
The low density polyethylene (LDPE) bag used to package implants adheres to the highly polished surface of the devices.
Edwards Lifesciences is recalling certain lots of EMBOL-X Glide Protection System due to deformation of the cannula tip.
HemoCue AB has become aware of a production related problem involving punctured HemoCue Hb 201 Microcuvettes single pack pouches.
The low density polyethylene (LDPE) bag used to package implants adheres to the highly polished surface of the devices.
The low density polyethylene (LDPE) bag used to package implants adheres to the highly polished surface of the devices.
The low density polyethylene (LDPE) bag used to package implants adheres to the highly polished surface of the devices.
MAQUET Getinge Group CARDIOHELP Base Unit REF 70104.7999 REF 70104.8012 Manufactured for: MAQUET Cardiopulmonary AG Made in Germany. Cardiopulmonary support system.
Maquet Medical Systems USA
It has come to the attention of MAQUET that in certain rare instances, the CARDIOHELP-i device may display an error message with audible alarm, indicating "battery defective." The alarm will occur-upon start up or when the main power supply is unplugged from the main power source. The human machine interface touch screen of the CARDIOHELP-i device may gradually become black due to the human machine interface (HMI) back light shutting down.
Dilators and Holding Clip are being used for a secondary function of neuromonitoring, for which it is not 510(k) cleared.
Xoran Minicat00202
Xoran Technologies
The MiniCAT CT Scanners subject to this notification failed to meet Xoran's specifications for dose rate of 125 ΒΏ 3 kVp. These 18 complaints received include dose rates both on the high and low end of Xoran's specifications. See ATTACHMENT 1 for details. (e) A n evaluation of the hazards reasonably related to defect or the failure to comply with the Federal standard; An evaluation of the health ha
Dilators and Holding Clip are being used for a secondary function of neuromonitoring, for which it is not 510(k) cleared.
An in-vitro diagnostic reagent may be labeled with an incorrect barcode, which when scanned on associated diagnostic instrument may cause a false negative result.
The recall was initiated after Varian Medical became aware the Lexan templates of the Breast Bridge Template System could become bent and turn opaque after steam sterilization.
One lot of 25 gauge Soft-Tip Disposable Cannulas are co-mingled with 23 gauge cannulas; the product is labeled as 25 gauge cannulas.
FSE Image Upgrade Kit, Catalog # 92019C This kit is used to install and/or upgrade software used with the EVOLIS Microplate System.
Bio-Rad Laboratories
The FSE Image Upgrade kit (part 92019C) is recalled due to issue related to the export of data from the EVOLIS analyzer to Laboratory Information Systems (LIS) when using the MONOLISA Anti-HBs EIA Quantitative Determination in EVOLIS APF version 6.0.
OPTI LION E-Plus Cassettes were labeled with an incorrect barcode. The incorrect barcode could perfect the performance of one or more parameters as the product ages. Also a Customer complaint from a distributor reported that a customer was getting QC failures.
Some devices may be contaminated on the back side of the device with glue in the blood application area.
Siemens determined that if an autosampler rack jam error occurs during operation of an ADVIA 2120/2120i connected to an ADVIA Autoslide, it is possible that the next slide processed by the Autoslide could be labeled with the wrong sample identification information.
GE Healthcare has recently become aware of a potential safety issue related with the DMPR function and Screen Saved Images on your scanner which involves Optima CT660, Brivo CT385, Discovery PET/CT 610, Discovery PET/CT 710. The DMPR and Screen Save functionality may possibly generate the same Series UID. It has been seen that some PACS may combine series under a patient using only the Series UID