Packaging error.
Medical Devices Recalls
Medical equipment, implants, and diagnostic devices
Arrow Three-Lumen CVC
ARROW INTERNATIONAL
Gravity flow rates on the affected product lidstock are incorrect.
When the TEG 5000 Analyzer including TEG Analytical Software is used with PlateletMapping (ADP or AA) Assays and TEG Mangement Software, incorrect values of the PlateletMapping ADP or AA % Inhibition and % Aggretagion may be displayed on TEG Manager for the affected assays.
Digital IVUS catheter instructions for use ( IFU) do not contain contrainidications
Backup batteries can fail earlier than their estimated life.
Medline Pill Splitters; 1 each/box; 144 boxes/carton
MEDLINE INDUSTRIES, LP - Northfield
Product lid (which houses a blade) is coming unhinged from the base, thus exposing the blade.
RayStation 11B
RAYSEARCH LABORATORIES AB
This notice concerns two issues found related to the display of Linear Energy Transfer (LET) in RayStation 11B including some service packs. First, when using a dose threshold for an evaluation dose, LET display might be misleading. Second, a displayed beam-specific LET distribution can sometimes be out of sync with the selected beam.
Sensica Urine Output System
C.R. Bard
Complaints relating to urine output measurement accuracy.
Device model and size for affected lots of Laser Fibers may be mismatched between the product's carton, immediate pouch packaging, and/or the device.
Mismatch between the expiration date on the sterile barrier pouch and the expiration date on the case/shipper label. Product sterile barrier pouch label expiration date of 16SEP2024 (2024-09-16) while the case and shipper labels both showed 21SEP2024 (2024-09-21), causing product being used for up to five (5) days past the true expiration date
EVIS EXERA II Gastrointestinal Videoscope-indicated for use within the upper digestive tract (including the esophagus, stomach, and duodenum). Model: GIF-HQ190
Olympus Corporation of the Americas
Part (A-rubber) intended for a different model of gastrointestinal videoscope may have been used when replacing part on GIF-HQ190 during the last repair at Olympus. An Incorrect A-rubber fits tighter on the GIF-HQ190 which would not meet the intended design specifications
BVI Wet -Field Eraser, 18GA Blunt Tip, 45Deg.-Intended to remove tissue and control bleeding by the use of high frequency electrical current Part Number: 221251
Beaver Visitec International
Boxes labeled as Wet-Field Eraser, 18GA Blunt Tip, 45Deg., were shipped containing incorrect product a Wet-Field Eraser, 18GA Bevel Tip, Straight, Part Number: 221250
Colonovideoscope Model CF-H180AL
Olympus Corporation of the Americas
A (Camera unit) CCD intended for a different model of colonoscope may have been installed in the CF-H180AL, viewing angle would be narrower than a normal CF-H180AL and the use of EndoTherapy devices may stick out (approximately 1.4 mm) further than normal protrusion,could lead to patient injury; specifically, mucosal damage that could result in hemorrhage or bleeding
REF: PIG1260K, 6Fr, X10, Safe-t-Centesis Catheter drainage Kit, Sterile EO,
Bard Peripheral Vascular
There is the potential that a 6 Fr Catheter Drainage kit contains an 8 Fr needle instead of a 6 Fr needle. They can be used in conjunction with the following drainage methods: Universal extension set and drainage bag, vacuum bottle, and wall suction.
Batteries for the HVAD system have a weld defect affecting internal components within the HVAD battery, resulting in the battery to malfunction and no longer provide power or prevent the battery from holding a complete charge or properly recharging.
Flow-c Anesthesia System : intended for use in administering anesthesia while controlling the entire ventilation of patient Product Code/REF Number: 6887700
Getinge Usa Sales
Potential for crack to form on the on/off switch of the suction unit, which may result in an unreasonable risk of harm to the patient. In the worst case, the switch can break, preventing the suction unit from being activated.
Mislabeled: Product labeled as Intergard Woven Straight graft REF IGW0032-30, D: 8 mm, L: 30 cm, contained an Intergard Woven Straight graft of D: 32 mm and L: 30 cm
Multiple issues with the potential for interruption of therapy or over-infusion: 1. Primary Audible Alarm (PAA), 2. Unanticipated Depleted Battery Alarms, 3. Time Base Alarm, 4. Intermittent Volume Over Time (IVOT) - Infusion Continues after System Failure, 5. Clearing of Program Volume Delivered (PVD), 6. False Alarm for Rate Below Recommended Minimum for Syringe Size, 7. Incorrect Bolus or Loading Dose Time Display, 8. Domain Name Server (DNS) Port 1001
Multiple issues with the potential for interruption of therapy or over-infusion: 1. Primary Audible Alarm (PAA), 2. Unanticipated Depleted Battery Alarms, 3. Time Base Alarm, 4. Intermittent Volume Over Time (IVOT) - Infusion Continues after System Failure, 5. Clearing of Program Volume Delivered (PVD), 6. False Alarm for Rate Below Recommended Minimum for Syringe Size, 7. Incorrect Bolus or Loading Dose Time Display, 8. Domain Name Server (DNS) Port 1001
Metalogix Slotted 90 Tubular Wrench x 10mm REF 900104 LOT A29618A
New Standard Device
Due to manufacturing processing error, wrench causing stripping when torqueing nuts.