Users need to be aware that embolism is a possible adverse event associated with hydrogel misplacements. Updated IFU provides new procedural instructions, warnings and precautions that describe steps to verify correct placement of the hydrogel following implantation, as well as technique recommendations for the proper placement of SpaceOAR and SpaceOAR Vue.
Medical Devices Recalls
Medical equipment, implants, and diagnostic devices
Acuson 9vE4 Transducer, Model (240)11289564 CE0129, Rx Only ,
Siemens Medical Solutions USA
Transducers (component of ultrasound system) may exceed recommended thermal index limits
Users need to be aware that embolism is a possible adverse event associated with hydrogel misplacements. Updated IFU provides new procedural instructions, warnings and precautions that describe steps to verify correct placement of the hydrogel following implantation, as well as technique recommendations for the proper placement of SpaceOAR and SpaceOAR Vue.
Smart Toe II Intramedullary Arthrodesis Implant
Stryker GmbH
The nickel-sensitivity statement on the outer carton may be partially covered by a label resulting in the statement showing "Indicated for patients with Nickel sensitivity" instead of "Not Indicated for patients with Nickel sensitivity".
A drug kit component was recalled by its manufacturer due to the potential for inadequacy of the validation of test methods used in its manufacture.
Drager Infinity CentralStation Software versions VG2.1.2 and VG2.1.2 SU12. For patient monitoring with Infinity M300 wireless telemetry devices.
Draegar Medical Systems
Software issue resulting in temporary loss of central monitoring functionality.
IMMULITE 2000 Systems HCG [200 Tests, material number 10381206, catalog number L2KCG] and [600 Tests, material number 10381194, catalog number L2KCG6]
Siemens Healthcare Diagnostics
Potential for falsely elevated hCG results due to sample carryover in serum and urine patient samples, as well as quality control samples and adjustors.
syngo.via RT Image Suite upgraded to VB60 running on SOMATOM go.Up Model Number 11061620. Running on servers with syngo.via VB60 software after upgrade and CT scanners with syngo.CT VA40A after upgrade.
Siemens Medical Solutions USA
Upgraded software version makes an automatic change in laser configuration settings for "markerless workflow" which may lead to the wrong laser offset coordinates being displayed in the Patient Marking step.
ZEISS miLOOP Lens Fragmentation Device REF 303071-9090-000. Used during cataract surgery.
Carl Zeiss Meditec AG
High friction of the slider can cause the device to stick, or not move as intended.
syngo.via RT Image Suite upgraded to VB60 running on syngo.via Model Number 10496180. Running on servers with syngo.via VB60 software after upgrade and CT scanners with syngo.CT VA40A after upgrade.
Siemens Medical Solutions USA
Upgraded software version makes an automatic change in laser configuration settings for "markerless workflow" which may lead to the wrong laser offset coordinates being displayed in the Patient Marking step.
A drug kit component was recalled by its manufacturer due to the potential for inadequacy of the validation of test methods used in its manufacture.
Puritan Bennett 980 Series Ventilator
Medtronic formerly Covidien
The audible alarm may not sound and/or the omni-directional LED visual alarm may not display during alarm states. Delayed awareness of alarm states may lead to delayed response or a delay of treatment, potentially resulting in hypoxia, dyspnea, cardiac arrest, or death.
Updated and revised user instructions issued 7/14/22. Potential carryover issue may cause elevated quality control and sample results for heparin, apixaban or rivaroxaban when HemosIL Liquid Anti-Xa reagent is used in the same run with both HemosIL Liquid Antithrombin and HemosIL Q.F.A. Thrombin (Bovine), specific to instrument models that utilize a single reagent probe
PrisMax System, Product Codes: a) 955558, b) 955627, c) 955701
Baxter Healthcare
This correction is due to software anomalies occurring during use. If the operator initiates therapy by entering a value other than the default value for the patient Gain/Loss Limit or the Return Pressure Drop Limit (RDL) and uses the Same Patient function, the PrisMax System will return the patient Gain/Loss Limit or the RDL to the default value, rather than the non-default value entered at the initiation of treatment. The PrisMax System interface will display the original values once treatment is restarted but does not have an alert to inform the user of the change. The RDL value will be returned to the original value immediately upon selecting the Same Patient function. The Gain/Loss Limit value will be returned to the original value when the patient weight and/or hematocrit is changed during treatment.
Empower Introducer Sheath with Hydrophilic Coating 21F x 56cm Model: 668782-200 (OUS)
Freudenberg Medical
Non-intended glue from the hub/shaft bond on the inside ID of the catheter shaft may cause Embolism leading to peripheral ischemia
PrisMax System, Product Code 955724
Baxter Healthcare
This correction is due to software anomalies occurring during use. If the operator initiates therapy by entering a value other than the default value for the patient Gain/Loss Limit or the Return Pressure Drop Limit (RDL) and uses the Same Patient function, the PrisMax System will return the patient Gain/Loss Limit or the RDL to the default value, rather than the non-default value entered at the initiation of treatment. The PrisMax System interface will display the original values once treatment is restarted but does not have an alert to inform the user of the change. The RDL value will be returned to the original value immediately upon selecting the Same Patient function. The Gain/Loss Limit value will be returned to the original value when the patient weight and/or hematocrit is changed during treatment.
There is a software anomaly with the InterStim X Clinician software application with results in the data entered in the Patient Information fields not e retained and a " Data Lost" notification being displayed.
WAVES WCM Kidney Cassette Module
Waters Medical Systems
There is a potential for the WCM cassette pump head leaking solution.
TearCare version 1.0 SmartHubs
Sight Sciences
Ophthalmic devices distributed prior to the recently-obtained 510(k) clearance do not fall within the relevant 510(k) exemption and so their performance characteristics were not evaluated by the FDA prior to distribution.
RocheDiabetes Care Platform Software version 2.5.2 material number 08888353001. This is a web application software hosted on a server. There is no physical product.
Roche Diabetes Care
Potential for patient data mismatch when using browser "back" button to navigate between patients when using the diabetes management software.