ðŸĨ

Medical Devices Recalls

Medical equipment, implants, and diagnostic devices

37,779 Total Recalls

Users need to be aware that embolism is a possible adverse event associated with hydrogel misplacements. Updated IFU provides new procedural instructions, warnings and precautions that describe steps to verify correct placement of the hydrogel following implantation, as well as technique recommendations for the proper placement of SpaceOAR and SpaceOAR Vue.

Feb 24, 2022 Infusion Pumps Nationwide View Details →

Users need to be aware that embolism is a possible adverse event associated with hydrogel misplacements. Updated IFU provides new procedural instructions, warnings and precautions that describe steps to verify correct placement of the hydrogel following implantation, as well as technique recommendations for the proper placement of SpaceOAR and SpaceOAR Vue.

Feb 24, 2022 Infusion Pumps Nationwide View Details →
Class I - Dangerous

The nickel-sensitivity statement on the outer carton may be partially covered by a label resulting in the statement showing "Indicated for patients with Nickel sensitivity" instead of "Not Indicated for patients with Nickel sensitivity".

Mar 28, 2022 Implants & Prosthetics Nationwide View Details →

Upgraded software version makes an automatic change in laser configuration settings for "markerless workflow" which may lead to the wrong laser offset coordinates being displayed in the Patient Marking step.

Mar 7, 2022 Diagnostic Equipment Nationwide View Details →

Upgraded software version makes an automatic change in laser configuration settings for "markerless workflow" which may lead to the wrong laser offset coordinates being displayed in the Patient Marking step.

Mar 7, 2022 Diagnostic Equipment Nationwide View Details →

Puritan Bennett 980 Series Ventilator

Medtronic formerly Covidien

Class I - Dangerous

The audible alarm may not sound and/or the omni-directional LED visual alarm may not display during alarm states. Delayed awareness of alarm states may lead to delayed response or a delay of treatment, potentially resulting in hypoxia, dyspnea, cardiac arrest, or death.

Mar 25, 2022 Patient Monitors Nationwide View Details →

Updated and revised user instructions issued 7/14/22. Potential carryover issue may cause elevated quality control and sample results for heparin, apixaban or rivaroxaban when HemosIL Liquid Anti-Xa reagent is used in the same run with both HemosIL Liquid Antithrombin and HemosIL Q.F.A. Thrombin (Bovine), specific to instrument models that utilize a single reagent probe

Feb 22, 2022 Other Medical Devices Nationwide View Details →

This correction is due to software anomalies occurring during use. If the operator initiates therapy by entering a value other than the default value for the patient Gain/Loss Limit or the Return Pressure Drop Limit (RDL) and uses the Same Patient function, the PrisMax System will return the patient Gain/Loss Limit or the RDL to the default value, rather than the non-default value entered at the initiation of treatment. The PrisMax System interface will display the original values once treatment is restarted but does not have an alert to inform the user of the change. The RDL value will be returned to the original value immediately upon selecting the Same Patient function. The Gain/Loss Limit value will be returned to the original value when the patient weight and/or hematocrit is changed during treatment.

Mar 15, 2022 Other Medical Devices Nationwide View Details →
Class I - Dangerous

This correction is due to software anomalies occurring during use. If the operator initiates therapy by entering a value other than the default value for the patient Gain/Loss Limit or the Return Pressure Drop Limit (RDL) and uses the Same Patient function, the PrisMax System will return the patient Gain/Loss Limit or the RDL to the default value, rather than the non-default value entered at the initiation of treatment. The PrisMax System interface will display the original values once treatment is restarted but does not have an alert to inform the user of the change. The RDL value will be returned to the original value immediately upon selecting the Same Patient function. The Gain/Loss Limit value will be returned to the original value when the patient weight and/or hematocrit is changed during treatment.

Mar 15, 2022 Other Medical Devices Nationwide View Details →

There is a software anomaly with the InterStim X Clinician software application with results in the data entered in the Patient Information fields not e retained and a " Data Lost" notification being displayed.

Feb 28, 2022 Other Medical Devices Nationwide View Details →
Class I - Dangerous

Ophthalmic devices distributed prior to the recently-obtained 510(k) clearance do not fall within the relevant 510(k) exemption and so their performance characteristics were not evaluated by the FDA prior to distribution.

Mar 23, 2022 Other Medical Devices Nationwide View Details →